The core idea
From heartbeat to map, in five links
Every mapping system — whatever the brand — is the same chain: sense the heart's local electrical signal, locate the electrode that sensed it in 3D space, and repeat that thousands of times until the points merge into a map. The differences between systems are in how they locate, how densely they sample, and how cleverly they interpret.
Electrogram
Electrodes on a catheter inside the chamber record the local activation signal — the electrogram (EGM) — at each spot they touch.
3D position
A magnetic field, impedance field, or dielectric field tells the system where each electrode is, several times per second.
Geometry
Sweeping the catheter over the chamber wall traces its shape, building a 3D shell of the atrium or ventricle.
Activation & voltage
Each point is coloured by when it fired (LAT) or how strong its signal was (voltage) — the map a clinician actually reads.
Ablation guidance
The map shows where the source or circuit lies. The clinical team uses it to guide ablation with RF, cryo or pulsed-field energy, and can map again to check the result.
Interactive · Try it
Activation map simulator
This is a schematic chamber, not a real heart — but the reading skill is the same. Pick a rhythm and watch how the wavefront moves. Red fires first, purple last. In a re-entrant rhythm the colours chase each other in a circle and never finish; in a focal rhythm they spread out from one early spot.
Focal atrial tachycardia
One irritable spot fires early and the wave spreads outward in rings — the map shows a single red “bullseye” with colours radiating away from it.
What to look for
Find the earliest (red) point first: in a focal rhythm it marks where activation begins, and everything else is the wave spreading from it.
Key numbers
Cycle length in this simulation ≈ 300 ms (illustrative)
In this model a single source drives the whole chamber.
Colour scale
Colours encode local activation time (LAT) — milliseconds relative to a reference electrogram, usually from a stable catheter in the coronary sinus.
In the EP lab
How a mapping case actually runs
Six stages, in order. Select each stage to see what the team is doing and what the mapping system contributes at that moment.
Explorer · Part 2 of 2 — The systems
The major mapping platforms, compared
This explorer compares five platforms described in published studies. They all build the same kinds of maps — what differs is the localisation physics, the signature catheter, and the design philosophy: closed and integrated, open to other brands' catheters, or built for maximum point density.
| System | Localisation | Signature mapping catheter | Known for |
|---|---|---|---|
| CARTO 3 | Magnetic sensor Gepstein 1997 + current-based (impedance) hybrid unverified | OCTARAY Sarkozy 2022 & PentaRay multi-spline | Integrated ecosystem, contact-force ablation, ICE integration unverified |
| EnSite X | Impedance (VoXel) + magnetic hybrid, open architecture unverified | Advisor HD Grid paddle Magtibay 2019 | Omnipolar mapping — voltage much less dependent on wavefront direction Deno 2017 |
| OPAL HDx (formerly Rhythmia HDx) unverified | Magnetic + impedance hybrid unverified | IntellaMap Orion mini-basket (64 electrodes) Anter 2015b | High-density automated mapping Sultan 2019 |
| KODEX-EPD | Dielectric sensing Abeln 2021 — no magnetic field generator unverified | Works with validated catheters from other makers unverified | Detailed anatomy imaging; studied in cryoballoon ablation Rottner 2021 |
| AcQMap | Ultrasound anatomy + non-contact charge-density sensing Pope 2022 | AcQMap balloon (48 ultrasound transducers, 48 electrodes) unverified | Whole-chamber activation, including during AF Pope 2022 |
Platform and catheter names are trademarks of their makers, used here only to identify the products. Facts carry a citation to a published study, listed at the bottom; items marked unverified have none yet. Manufacturer pages are listed for further reading, not as sources.
Reading the map
One chamber, many maps
The geometry stays the same; what changes is the value painted onto it. Choosing the right map type is most of the diagnostic skill.
Signal lab
Reading an electrogram
The map is only as good as the signals underneath it. These schematic traces show the four electrogram shapes a mapper learns to recognise on sight. Amplitude and timing — not the picture — are what the system measures.
Hardware
The catheters that do the sensing
High-density mapping became possible when catheters grew from a few electrodes to dozens of small, closely spaced ones. Smaller electrodes with closer spacing record more local signals Anter 2015a.
Multi-spline
PentaRay & OCTARAY (CARTO)
Flexible splines that splay against the chamber wall like a star — five splines carrying 20 electrodes on the PentaRay unverified, eight splines carrying 48 on the OCTARAY Masuda 2024 — giving a wide footprint of points per placement.
Grid paddle
Advisor HD Grid (EnSite)
A flat 4 × 4 paddle of 16 electrodes unverified. Because it samples along two axes at once, the system can compute omnipolar signals Magtibay 2019 — an amplitude that depends much less on the direction the wave crosses the paddle (course module m23).
Mini-basket
IntellaMap Orion (OPAL HDx)
An expandable basket with 64 small electrodes Anter 2015b on 8 splines unverified. Paired with an automated mapping system Sultan 2019, it collects high-density maps — the platform's design goal.
Circular
Lasso-style circular catheters
A loop of electrodes that sits at a pulmonary vein opening and records the vein's potentials — used to check pulmonary vein isolation during AF ablation Anter 2015b.
Ablation
Contact-force ablation catheters
The treatment catheter (e.g. ThermoCool SmartTouch Teijeira-Fernandez 2018; TactiFlex and IntellaNav StablePoint unverified) senses tip contact force in grams, so lesion tags on the map reflect pressure, power and time — not just position. Contact also changes the voltage recorded Teijeira-Fernandez 2018.
Non-contact
AcQMap balloon
A balloon carrying ultrasound transducers and electrodes that hovers in the chamber without touching the wall, reconstructing charge density over the whole chamber at once Pope 2022.
Reference
Glossary
The vocabulary of the EP lab, in plain language. Search any term.
Check yourself
Which map answers the question?
Four short scenarios. Answer all four and get a read on where your mapping instincts are strong — and what to study next.
Keep going
A learning roadmap
For a student, trainee, engineer or curious builder — the order that makes the field click.
Heart electricity basics
Before any technology:
- Action potentials & conduction
- The ECG and what each wave means
- Atria, ventricles & the conduction system
Arrhythmia mechanisms
The three ways rhythms go wrong:
- Enhanced automaticity (focal)
- Re-entry circuits
- Triggered activity
Signals & maps
The mapper's core craft:
- Bipolar vs unipolar EGMs
- LAT annotation & windows of interest
- Voltage thresholds & scar
Systems & cases
Put it together:
- Compare the five platforms above
- Follow published case reports
- Entrainment & ablation endpoints (PVI, block lines)
Map-reading checklist
- Reference stable? — LAT is only meaningful against a steady reference signal
- Window set right? — in re-entry, maps degrade when the window of interest is shorter than the cycle length Koa-Wing 2015
- Early meets late? — in re-entry, red should sit next to purple around the circuit
- Density sufficient? — gaps in points are guesses, not data
- Voltage direction-checked? — a wave crossing at right angles to a bipole reads falsely low
- Scar or artefact? — low voltage can also come from poor contact Teijeira-Fernandez 2018 or wave direction
- EGMs inspected? — the colour is only as good as the signals behind it