Glossary
Terms used in the course, in alphabetical order. Each entry names the modules that use it. Clinical terms are defined in plain words for engineers; they are not clinical guidance.
- Ablation
- Destroying a small area of heart tissue on purpose, for example by radiofrequency heating, to stop an arrhythmia. m03
- Action potential
- The rise and fall of the transmembrane voltage when a heart cell is excited. m03, m08
- Action potential duration (APD)
- How long a cell stays depolarized after an upstroke, measured at a stated repolarization level. m08
- Activation map
- A map colored by local activation time, interpolated between the mapped points. m24
- Activation wavefront
- The moving boundary between excited and resting tissue. Also called the front. m03
- Active catheter tracking
- Locating small receiver coils at a catheter’s tip from MR signals so the scanner can follow it and steer the imaging plane; reported at up to 50 updates per second. m53
- Adversarial example
- An input changed slightly, often invisibly, so that a model gives a different output; for ECG models, a small added waveform. m46
- AH interval
- Time from the atrial to the His deflection on the His-bundle recording; mostly conduction through the AV node. m11
- AI agent
- A program in which a model, usually a large language model, chooses step by step which tool to call, reads the result and continues until a goal is met or a step limit is reached. m44
- Air kerma
- The kinetic energy that X-rays release per unit mass of air, in gray (Gy); measured at a reference point it relates to the dose at the patient’s skin. m49
- Aliasing
- The folding of frequencies above half the sampling rate onto lower ones; once sampled, the two cannot be told apart. m04
- Alternans
- Beat-to-beat alternation of action potential duration (or another quantity) at a steady pacing rate. m10
- Analytic signal
- A complex signal whose real part is the original and whose imaginary part is its Hilbert transform; its angle gives an instantaneous phase, meaningful when the signal is narrowband. m05
- Anisotropy
- Direction dependence of a tissue property: heart muscle conducts faster, and has lower resistance, along its fibers than across them. m01, m09
- Applied part
- The part of medical electrical equipment that contacts the patient in normal use; IEC 60601-1 classes include CF (cardiac floating) for direct cardiac connection. m15
- Arrhythmia
- An abnormal heart rhythm. m03
- Atrial fibrillation (AF)
- A fast, irregular rhythm of the upper chambers, without organized atrial activity. m37, m10
- Atrium (plural atria)
- One of the two upper chambers of the heart. m03
- Attention
- In a neural network, a weighted average of value vectors whose weights come from how well a query matches each key, normalized by a softmax. The building block of the transformer. m43
- Audit log
- An append-only record of every step an agent took: the tool, its arguments, the policy decision and the result, so a person can check the run afterwards. m44
- Augmented limb leads
- aVR, aVL and aVF: each limb electrode measured against the mean of the other two. m12
- Average precision (PR AUC)
- The area under the precision–recall curve, averaged over thresholds. Unlike the ROC AUC it falls when positives are rare. m39
- Average symmetric surface distance (ASSD)
- The mean distance between two surfaces, measured from each to the other; a typical-error measure for segmentations. m40
- Basket catheter
- A catheter whose electrodes sit on flexible splines that open into a cage filling a chamber, for simultaneous recording over a wide area. m13
- Bayesian optimization
- Choosing where to measure next by modelling an unknown function with a probabilistic surrogate, such as a Gaussian process, and maximizing an acquisition function. m41
- Bidomain model
- A tissue model with separate intracellular and extracellular potentials and conductivities, coupled through the membrane. m09
- Bipolar electrogram
- The difference between the signals of two nearby electrodes. It acts like a spatial derivative: it keeps local detail and rejects distant activity, but depends on the bipole’s orientation to the wavefront. m03, m13, m23
- C-arm
- A fluoroscopy system with the X-ray source and detector on opposite ends of a C-shaped arm that rotates around the patient. m55
- Cable equation
- The equation for voltage along a passive fiber with leaky membrane and axial resistance. m09
- Cardiac sarcoidosis
- Inflammation of the heart by granulomas, which can cause conduction block and ventricular tachycardia; FDG PET can show active inflammation. m54
- Cavotricuspid isthmus (CTI)
- The strip of right atrial tissue between the inferior vena cava and the tricuspid valve; the usual ablation target for typical atrial flutter. m50
- Central-slice theorem
- The 1-D Fourier transform of a projection equals the 2-D Fourier transform of the image along the line through the origin at the same angle; it is why projections at enough angles determine a CT slice. m51
- Clique
- A small group of electrodes, such as a triangle or a square, whose bipoles are combined to estimate the local electric field. m03, m25
- Cohen’s kappa
- Agreement between two raters or methods beyond what chance would give: (observed − expected agreement) / (1 − expected agreement). m53
- Collocation point
- A point where a governing equation is checked during fitting; it needs no measurement. m42
- Common-mode rejection ratio (CMRR)
- The ratio of an amplifier’s gain for the difference between its inputs to its gain for their mean, often in decibels. m14
- Common-mode voltage
- The mean of the voltages at a differential amplifier’s two inputs; on the body it is mostly picked up from the mains. m14
- Complex conductivity
- Conductivity plus the displacement-current term, sigma + j omega epsilon0 epsilon_r; its phase grows with frequency and differs between tissues. m18
- Computed tomography (CT)
- Imaging that measures X-ray attenuation along many lines through the body and reconstructs cross-sectional slices from those measurements. m51
- Concealed conduction
- An impulse that is blocked but still resets the excitability of the tissue it entered, so it affects the next beat. m11
- Concealed fusion
- Entrainment of a tachycardia with no change in the surface QRS (or P wave) morphology during pacing. m11
- Concordance index (C index)
- For a risk score, the share of comparable patient pairs in which the patient with the earlier event had the higher score; 0.5 is chance. m40
- Condition number
- The ratio of a matrix’s largest to smallest singular value; it bounds how much a relative error in the data can grow in the solution. m07, m13
- Conducting channel
- A corridor of surviving tissue through scar that may carry a reentrant ventricular tachycardia; post-processing software can identify candidate channels on LGE. m52
- Conduction velocity
- The speed at which the activation wavefront travels through tissue, in m/s (1 m/s = 1 mm/ms). m03, m42, m06, m09, m25
- Conformal prediction
- A method that turns a model’s scores into a set of possible answers that contains the true one at a chosen rate, using a held-out calibration set. m46
- Contact force
- The force between a catheter tip and the tissue, measured from the deformation of a small spring element near the tip. m13, m23
- Contrast-to-noise ratio (CNR)
- The difference between two tissues’ mean signals divided by the noise standard deviation; it decides whether they can be told apart, and averaging n independent samples multiplies it by the square root of n. m48, m49
- Contrastive learning
- Self-supervised learning that pulls the embeddings of matched pairs, such as two recordings of one patient, together and pushes unmatched pairs apart. m43
- Convex hull
- The smallest convex set containing a set of points; it cannot follow concavities. m20
- Core-conductor model
- A model of a cell strand in which intracellular current flows along the strand and leaks out through the membrane; it gives membrane current as the second derivative of transmembrane voltage. m01
- Cotangent Laplacian
- The standard discretization of the Laplace–Beltrami operator on a triangle mesh: each edge is weighted by half the sum of the cotangents of the two angles opposite it. m06
- Cross-entropy
- The loss most classifiers minimize: minus the log of the probability given to the true class. m37
- Current dipole
- A current source and an equal sink very close together; its strength is current times separation, and its potential falls as one over distance squared. m01
- CUSUM
- Cumulative sum: a running sum of deviations above a target, reset at zero, that raises an alarm when it crosses a threshold. Used to detect a lasting shift in a monitored quantity. m45
- Cycle length
- The time between successive activations in a repeating rhythm, in ms. m22
- Data augmentation
- Training on altered copies of the data (for example, flipped or noisier signals) so a model ignores the alteration. m37
- Dataset shift
- Any difference between the data a model was trained on and the data where it is used: in class mix (prior shift), in the recordings (covariate shift) or in what labels mean. m37
- De-identification
- Removing or coarsening information in health data so that it no longer identifies a person, for example by removing direct identifiers and grouping quasi-identifiers; the rules differ by country. m47
- Decapolar catheter
- A catheter with ten electrodes along its shaft. m03, m39
- Deceleration zone
- A region of crowded isochrones on an activation map, defined in one study as more than 3 isochrones within a 1-cm radius. m24
- Diastolic interval (DI)
- The time from the end of one action potential to the start of the next. m08
- Dice coefficient
- Overlap of a predicted and a true region: twice the shared area over the sum of the two areas, from 0 (no overlap) to 1 (identical). m39, m40, m21
- Dielectric imaging
- Building anatomy and tissue information from how electrical fields are shaped by the conductivity and permittivity of what surrounds the electrodes. m18
- Discrepancy principle
- Choosing a regularization parameter so that the model fits the data only as closely as the noise level allows. m07, m32
- Discrete Picard condition
- The requirement that the data’s coefficients in the singular vectors fall faster than the singular values; without it no regularized solution is good. m07
- Dominant frequency
- The frequency of the largest spectral peak in a search band; for electrograms it tracks the activation rate only after suitable preprocessing. m05, m29
- Double layer
- A surface covered with current dipoles pointing normal to it; a uniform double layer’s potential is proportional to the solid angle it subtends. m01
- Drift
- Slow change of reported positions over a procedure, for example from changing chest impedance; estimated by revisiting landmarks. m19
- Driven right leg
- A circuit that senses the body’s common-mode voltage, inverts and amplifies it, and drives it back through the right-leg electrode to reduce it. m14
- Duty cycle
- The fraction of time a gate accepts data. m19
- Early meets late
- In an activation map of a reentry, the place where the earliest and latest times sit side by side; it marks where the window of interest wraps, not necessarily a block line. m22
- ECG gating
- Using the electrocardiogram to keep only imaging data from a chosen part of the cardiac cycle, so that the moving heart appears still. m51
- Eddy currents
- Currents induced in a conductor by a changing magnetic field; their own field distorts magnetic localization. m16
- Edema (myocardial)
- Excess water in heart tissue after injury such as ablation; bright on T2-weighted MRI, it includes reversible injury and fades over time. m53
- Effective refractory period (ERP)
- The shortest coupling interval at which a premature stimulus still captures the tissue. m08
- Eikonal equation
- The equation v |grad T| = 1 linking an activation-time map T to the local conduction speed v. m42, m06
- Einthoven triangle
- The idealization in which the three limb leads have lead vectors 60 degrees apart around a centric heart vector. m12
- Electrical remodeling
- Changes in ion channels and action potentials caused by an arrhythmia itself, such as shorter atrial refractoriness during atrial fibrillation. m08
- Electrogram
- The voltage an electrode records from currents crossing cell membranes nearby. m03
- Embedding
- A vector that represents an input, such as a beat or a sentence, so that similar inputs get nearby vectors. m43
- Endocardium
- The inner surface of the heart wall. m03
- Ensemble
- A set of models trained with different random choices; their spread is used as an uncertainty estimate. m42
- Entrainment
- Capture of a reentrant circuit by pacing slightly faster than the tachycardia, without stopping it. m11
- Error budget
- A list of the error sources in a chain of steps with their sizes; independent, unbiased errors add in quadrature, and biases add directly. m55, m26
- Euler characteristic
- V − E + F for a mesh with V vertices, E edges and F faces; 2 for a closed surface without handles, one less for each hole cut into it. m06
- Excitability
- The ability of a membrane to respond to a stimulus above threshold with a full action potential. m08
- Excitable gap
- The recovered tissue ahead of a reentrant front: circuit length minus wavelength, or cycle length minus refractory period. m10
- Expected improvement
- An acquisition function: the average amount by which a new measurement is expected to beat the best value so far. m41
- External validation
- Testing a frozen model on data from another place or time that played no part in training or tuning. m37
- Extracellular volume (ECV)
- The fraction of a tissue’s volume outside cells. MRI estimates it from T1 before and after gadolinium and the hematocrit; scar and fibrosis raise it. m52
- Far-field signal
- Activity that an electrode picks up from tissue that is not directly under it. m03, m01, m22
- Fast marching method
- An algorithm that solves the eikonal equation by advancing a front from the source in order of arrival, on grids or triangle meshes. m06
- FDG (fluorodeoxyglucose)
- A glucose analog labeled with fluorine-18; cells that use much glucose, including inflammatory cells, take it up and show on PET. m54
- Feature attribution
- A score for how much each input feature, such as each sample of an ECG, contributed to a model’s output for one input. m46
- Federated learning
- Training one model across sites that keep their data; only model updates are shared. m37
- Fibrosis
- Replacement of heart muscle by connective tissue. In the atria it can be estimated on late-gadolinium MRI. m40
- Fiducial localization error (FLE)
- The error in locating a landmark used for registration, in either the image or the map. m55, m21
- Fiducial registration error (FRE)
- The residual distance between registered landmarks after a fit; its expected value depends on the number of landmarks and FLE, not on their layout. m55, m21
- Field generator
- The set of coils of known position and orientation that a magnetic localization system drives to create its reference fields. m16
- Field scaling
- Calibrating an impedance-based localization field with electrodes of known spacing, so measured potential differences give true distances. m17
- Fill distance
- The largest distance from any point of a surface to its nearest sample point or electrode; a measure of coverage. m13
- Fill threshold
- The distance from the nearest mapped point beyond which a mapping system leaves the surface uncolored. m24
- Filter factor
- The weight a regularization method gives each singular component of the solution: 1 or 0 for truncated SVD, s²/(s² + λ²) for Tikhonov. m07
- Filtered back-projection
- CT reconstruction that filters each projection with a ramp in frequency and smears it back along its lines; without the filter the image is blurred. m51
- FitzHugh–Nagumo model
- A two-variable model of an excitable membrane, with a fast voltage-like variable and a slow recovery variable. m08
- Fluoroscopy
- Continuous or pulsed X-ray imaging that shows a moving shadow image, used to guide catheters, sheaths and leads. m49
- Focal loss
- A cross-entropy that down-weights examples the model already gets right, so training attends to rare and hard ones. m39
- Focal source
- A small region from which activation spreads outward. m03, m10
- Forward problem
- Computing body-surface (or electrode) potentials from known cardiac sources and a model of the torso. m12
- Fossa ovalis
- The thin central part of the wall between the two atria, where a needle usually crosses from the right atrium to the left. m50
- Foundation model
- A large model pretrained on broad, mostly unlabeled data and then adapted to many downstream tasks. m43
- Frontal axis
- The direction of the heart vector in the frontal plane, usually from leads I and aVF; 0 degrees toward the left arm, +90 toward the feet. m12
- Full width at half maximum (FWHM)
- The width of a blur between the two points where it falls to half its peak; for a Gaussian, about 2.355 standard deviations. Used to quote the spatial resolution of scanners. m48, m51, m54
- Gadolinium contrast
- An MRI contrast agent that shortens T1 in proportion to its concentration and stays outside cells, so it accumulates where there is more extracellular space. m52
- Gating variable
- A number between 0 and 1 for the open fraction of one kind of channel gate, opening and closing at voltage-dependent rates. m08
- Gaussian mixture model
- A probability model that describes data as a weighted sum of normal distributions; fitted with the EM algorithm to find clusters without labels. m39
- Gaussian process
- A probability model over functions; given measurements, it gives a mean estimate and an uncertainty at every other point. m41, m26
- Generalized cross-validation (GCV)
- Choosing a regularization parameter by minimizing a predicted-error function of the data, without a noise estimate. m07
- Geodesic distance
- The length of the shortest path between two points that stays on a surface. m41, m06
- Graph convolution
- A network layer that mixes each point’s features with those of its neighbors on a graph, such as the vertices of a mapping mesh. m39
- Grating lobe
- A false, full-strength copy of an array’s beam that appears when its elements are spaced more than half a wavelength apart for the steering used; echoes from it are drawn in the wrong direction. m50
- Group delay
- How long a filter delays the envelope of a signal at a given frequency; causal filters shift features such as activation times by it. m04
- Guardrail
- A rule enforced outside the model that limits what an agent may do, such as a tool allowlist, a schema check or a human approval step. m44
- Half-life
- The time for a radioactive tracer’s activity to halve; 109.73 minutes for fluorine-18. m54
- Half-value layer (HVL)
- The thickness of a material that halves the intensity of an X-ray beam; for a narrow beam of one energy it is ln 2 divided by the attenuation coefficient. m49
- Hallucination
- Output of a language model that reads as fluent and confident but is false or unsupported, such as an invented number or reference. m43
- Harmonic interpolation
- Filling in values on a mesh by solving Laplace’s equation between the known vertices; it cannot overshoot the data. m24
- Hausdorff distance
- The largest distance from any point of one surface to the other surface, in either direction; one stray island sets it. m40, m20
- Heart vector
- The single equivalent dipole that sums the heart’s current sources at one instant. m12
- Heat method
- A way to compute geodesic distance on a mesh: diffuse heat briefly from the source, normalize its gradient, then solve a Poisson equation. m06
- Hematocrit
- The fraction of blood volume made up of red cells; gadolinium in blood is confined to the rest, the plasma. m52
- High-density grid
- A catheter whose electrodes form a planar array, so bipoles can be formed along two perpendicular directions. m13
- His bundle
- The bundle of conducting fibers that carries activation from the AV node to the bundle branches and Purkinje system. m11
- Hopf bifurcation
- The point at which a steady state loses stability and small oscillations appear; in excitable cells, the onset of repetitive firing. m08
- Hounsfield unit (HU)
- The CT number scale, 1000 (μ − μ_water)/(μ_water − μ_air): water is 0 and air −1000. Values depend on the X-ray beam’s energy. m51
- Human approval step
- A point in an agent’s run where a person must accept or decline an action, such as any write, before it is carried out. m44
- HV interval
- Time from the His deflection to the earliest ventricular activation; conduction through the His–Purkinje system. m11
- Hybrid localization
- Correcting impedance-based electrode positions with points where a magnetic sensor gives the true position, so every electrode can be shown accurately. m18
- Ill-posed problem
- A problem whose solution fails to exist, to be unique or to depend continuously on the data; small data errors then cause large solution errors. m07
- Image integration
- Bringing a CT or MRI anatomy into a mapping system’s coordinates by segmentation and registration, so the catheter can be shown in it. m21
- Image intensity ratio (IIR)
- An LGE wall intensity divided by the mean blood-pool intensity, used to compare scans that have no absolute intensity scale; thresholds on it define atrial fibrosis or scar. m52
- Impedance localization
- Locating catheter electrodes from the potentials that high-frequency currents between skin patches make at them. m17
- Implantable loop recorder (ILR)
- A small device placed under the skin that records the ECG for years and sends alerts when its detector fires. m38
- Instrumentation amplifier
- A differential amplifier with a high-impedance, gain-setting input stage ahead of a difference stage, which multiplies its common-mode rejection. m14
- Integrated gradients
- A feature-attribution method that averages the model’s gradient along a straight path from a baseline input to the actual input; the attributions add up to the change in output. m46
- Interventional MRI
- Performing a procedure, such as catheter ablation, inside an MRI scanner with real-time images instead of fluoroscopy. m53
- Intracardiac echocardiography (ICE)
- Ultrasound imaging from a catheter inside the heart, used during procedures to see anatomy and build shells. m40
- Intrinsic deflection
- The steepest downstroke of a unipolar electrogram, used to mark local activation. m03, m22
- Inverse problem
- Recovering sources from measurements, the reverse of the forward problem that predicts measurements from sources. m07, m32
- Inverse-distance weighting (IDW)
- Interpolation by a weighted mean of the data with weights proportional to distance to the power -p; for p > 1 it is flat at each data point. m24
- Inverse-variance weighting
- Combining independent unbiased estimates with weights proportional to one over their variances; the result is at least as precise as the best one. m18
- Inversion recovery
- An MRI preparation that flips the longitudinal magnetization with a 180° pulse and reads the image while it recovers; a tissue has no signal at T1 ln 2. m52
- Iodinated contrast
- An iodine-containing agent injected into the blood; iodine attenuates X-rays strongly, so the blood pool stands out from the heart wall on CT. m51
- Isocenter
- The point a C-arm rotates about; structures there are imaged at a fixed magnification. m55
- Isochrone
- A line joining sites activated at the same time; isochrones a fixed time step apart are closer where conduction is slower. m24
- Isolation barrier
- The insulation and spacing that separate patient-connected circuits from mains-powered circuits; its capacitance still passes a small leakage current. m15
- Isthmus (VT)
- A narrow channel of conducting tissue, bounded by scar or anatomical barriers, that a reentrant ventricular tachycardia passes through. m51
- Iterative closest point (ICP)
- Surface registration that alternately pairs each point with its nearest point on the other surface and fits a rigid transform to those pairs; it converges to the nearest local minimum. m21
- Jaccard index
- Overlap of two regions as intersection over union; Dice = 2J/(1 + J). m40, m21
- k-anonymity
- A property of a released table: every record shares its quasi-identifier values with at least k − 1 others, so a match by those values points to at least k records. m47
- Kerma-area product (KAP)
- Air kerma times the area of the X-ray beam, in Gy·cm²; the same at any distance along the beam, so it describes the total radiation delivered. Also called dose-area product. m49
- L-curve
- A log-log plot of solution size against residual size over all regularization parameters; its corner is one way to choose the parameter. m07, m32
- Landmark registration
- Registration from paired points located in both coordinate systems, solved in closed form by least squares with an SVD. m21
- Laplace–Beltrami operator
- The Laplacian on a curved surface: the divergence of the surface gradient. m06
- Large language model (LLM)
- A transformer trained on very large amounts of text to predict the next token; it can follow instructions and answer questions in text. m43
- Larmor frequency
- The frequency at which nuclear magnetization precesses in a magnetic field: 42.58 MHz per tesla for protons, about 64 MHz at 1.5 T. m52
- Late gadolinium enhancement (LGE)
- MRI taken some minutes after a gadolinium contrast injection, when scar and fibrosis appear bright. m40, m52
- Lead field
- An electrode’s sensitivity to sources everywhere: by reciprocity, the gradient of the potential it would create by injecting unit current. m02, m12
- Lead vector
- The vector L for which a lead’s voltage equals L . p for a dipole p; it depends on the electrodes and the torso. m12
- Leakage current
- Current that flows from equipment through an unintended path, such as the patient to ground, in normal use or with a single fault. m15
- Leave-one-out cross-validation
- Scoring a model by predicting each data point from all the others; for a Gaussian process it needs only one matrix inverse. m26
- Leave-one-patient-out
- A test scheme that trains on all patients but one and tests on that one, in turn, so no patient is in both sets. m39
- Left atrial appendage occlusion (LAAO)
- A catheter procedure that closes the left atrial appendage with an implanted device. m50
- Length constant
- The distance over which a held voltage decays to 1/e along a passive cable, the square root of membrane over axial resistance. m09
- Length scale (kernel)
- The distance over which a Gaussian process’s covariance falls off, which sets how far each point’s influence reaches. m26
- Lesion gap
- A segment of an intended ablation line where the tissue was not destroyed, through which conduction can recover. m53
- Lesion metabolic activity
- A PET measure of a lesion: its partial-volume-corrected mean uptake times its volume. m54
- Levenberg–Marquardt method
- An iterative method for nonlinear least squares that blends Gauss–Newton steps with damped gradient steps. m42, m16
- Likelihood ratio
- How much a test result changes the odds of disease: for a positive result, sensitivity divided by one minus specificity; for a negative result, one minus sensitivity divided by specificity. Post-test odds are pre-test odds times the likelihood ratio. m48
- Line of block
- A line across which activation does not pass directly, so sites on either side activate at very different times. m24
- Linear probe
- A linear classifier trained on the frozen features of a pretrained model, used to measure how much those features already encode. m43
- Local activation time
- The moment the activation wavefront reaches a given site. m03, m22
- Local plane fit
- Estimating the activation-time gradient by fitting a plane to the activation times of nearby points by least squares; more points reduce noise, a larger radius adds bias where activation curves. m25
- Locally Delaunay edge
- A mesh edge whose two opposite angles add up to at most 180°; its cotangent weight is then not negative. m06
- Low-voltage area
- A region whose bipolar voltage falls below a chosen cut-off. It is often read as scar, but the voltage also depends on the catheter and the wavefront. m03, m23
- Macro-reentry
- Reentry around a large circuit set by anatomy or scar, such as typical atrial flutter. m10
- Macroshock
- Electric shock from current entering and leaving through the skin, so it spreads through the body before reaching the heart. m15
- Magnetic dipole
- The far-field model of a small current loop, with moment m; its field falls as the cube of distance. m16
- Magnetic resonance imaging (MRI)
- Imaging of hydrogen nuclei in a strong magnetic field, with contrast from how their magnetization relaxes and position encoded by field gradients. m52
- Magnetohydrodynamic voltage
- A voltage induced when conducting blood flows across a strong magnetic field; it distorts the ECG and electrograms inside an MRI scanner. m53
- Magnification (projection)
- How much larger a structure appears on the detector than its true size: source-to-detector over source-to-object distance. m55
- Marching cubes
- An algorithm that extracts a triangle surface at a chosen level from volumetric data. m20
- Membrane current
- Current crossing the cell membrane, positive outward; ahead of a wavefront it flows out (a source) and behind it flows in (a sink). m01
- Method of images
- Solving a problem with a flat boundary by adding mirror sources that make the boundary conditions hold; for two conductors the mirror source has strength (sigma1 - sigma2)/(sigma1 + sigma2) times the original. m02
- Microshock
- Electric shock from current delivered directly to the heart, for example through a catheter, where microamperes can capture or fibrillate the ventricles. m15
- Mitchell–Schaeffer model
- A two-variable cardiac membrane model with one gated inward current and one outward current; its action potential duration has a closed form. m08
- Model card
- A short document released with a trained model that states its intended use, out-of-scope uses, training and evaluation data, metrics overall and by group, and limitations. m47
- Monodomain model
- A tissue model with one potential and one effective conductivity; exact in 1D or when both domains are equally anisotropic. m09
- Morphological closing
- Dilation followed by erosion with the same shape; it fills gaps and concavities narrower than the shape while keeping larger features. m20
- Myocardial infarction
- Death of an area of heart muscle caused by loss of its blood supply. m03
- Negative predictive value (NPV)
- The share of negative results that are truly negative; like the positive predictive value, it depends on prevalence. m48, m35
- Nernst potential
- The voltage at which an ion’s diffusion and electrical drift balance: (RT/zF) ln of the outside-to-inside concentration ratio. m08
- Net benefit
- In decision-curve analysis, true positives minus false positives weighted by the odds of the threshold probability, per patient. It compares a model with acting on everyone or no one. m45
- Noise factor
- Total input-referred noise power divided by the noise power of the source resistance alone; it is lowest when the source resistance equals the amplifier’s voltage noise over its current noise. m14
- Noise floor
- The smallest amplitude a measurement can show: with no signal, the peak-to-peak of noise in the measurement window. m23
- Noncontact mapping
- Reconstructing endocardial potentials from a multielectrode probe that floats in the cavity, by solving an inverse problem. m07
- Normalized surface Dice (NSD)
- The share of two surfaces lying within a chosen tolerance of each other. m40
- Notch filter
- A narrow band-stop filter used to remove mains interference at 50 or 60 Hz. m04
- Nyquist frequency
- Half the sampling rate: the highest frequency a sampled signal can represent without aliasing. m04
- Omnipolar electrogram
- A signal built from a clique by estimating the local electric field and reading it along the direction of propagation, which removes most orientation dependence. m03, m23, m25
- Otsu’s threshold
- The intensity cut that splits a histogram into two classes with the largest between-class variance; used as a data-driven threshold for scar. m52, m21
- Out-of-distribution (OOD) input
- An input unlike the data a model was trained on, such as a recording from a new device or with heavy noise, for which its outputs are not reliable. m46
- Outflow tract
- The part of a ventricle leading to its great artery (pulmonary artery on the right, aorta on the left); a common origin of idiopathic premature ventricular beats. m41
- Oversampling
- Sampling faster than needed and averaging; for white noise each doubling of the averaged samples gains about 3 dB, half a bit. m14
- Pace mapping
- Pacing from a catheter at candidate sites and comparing the paced 12-lead ECG with the ECG of the arrhythmia, to find where the arrhythmia starts. m41
- Pacing
- Stimulating the heart electrically from an electrode at a chosen site. m03
- Partial-volume effect
- A voxel, or a region blurred by the scanner, that holds more than one tissue reads a mixture of their values, so thin or small structures lose contrast and change apparent size. m51, m54, m48
- Passive catheter visualization
- Seeing a catheter on real-time MR images through the signal void or artifact its materials cause; it shows one plane at a time. m53
- Patch pair
- Two skin electrodes that drive a localization current through the chest; three roughly orthogonal pairs give three coordinates. m17
- Peak-to-peak amplitude
- The difference between the largest and smallest values of a signal in a window; the usual electrogram voltage. m23
- Phase singularity
- A point around which the phase of activation takes every value; a spiral wave turns around one. m10
- Phase-sensitive inversion recovery (PSIR)
- An LGE reconstruction that keeps the sign of the recovering magnetization, so it is much less sensitive to the chosen inversion time than a magnitude image. m52
- Phased array
- A row of transducer elements whose firing delays steer and focus the beam electronically, without moving the probe. m50
- Photoplethysmography (PPG)
- Measuring blood-volume changes in tissue with light; the basis of pulse sensing in watches and trackers. m38
- Physics-informed neural network (PINN)
- A neural network trained on a loss that adds the residual of a governing equation to the data misfit. m42
- Point-spread function (PSF)
- The image an imaging system makes of a single point: the blur it applies to everything it shows. m48
- Poisson’s equation
- The equation the potential obeys in a volume conductor, div(sigma grad phi) = -I_v: Ohm’s law plus current conservation. m02
- Pose
- Position together with orientation; a single sensor coil gives five of its six degrees of freedom. m16
- Positive predictive value (PPV)
- The share of positive results (alerts, flags) that are truly positive; it depends on prevalence. m37, m38, m48
- Positron emission tomography (PET)
- Imaging of a tracer that emits positrons; each annihilation sends two photons in opposite directions, detected together to define a line through the source. m54
- Post-pacing interval (PPI)
- Time from the last pacing stimulus to the next local activation at the pacing site after entrainment. m11
- Pre-test probability
- The probability of a condition before a test result is known, for example its prevalence in the group tested. m48
- Predetermined change control plan (PCCP)
- A plan, authorized with an AI-enabled device’s marketing submission, that describes planned changes to the device and how each will be validated, so those changes need no new submission. m45
- Premature beat
- A beat that comes earlier than expected, usually followed by a longer interval; a common cause of false AF alerts. m38
- Premature ventricular contraction (PVC)
- An extra beat that starts in a ventricle earlier than the next normal beat. m41
- Prompt injection
- Text inside a document or tool output that tries to make a language model follow new instructions, such as a hidden request to delete files. m44
- Pulmonary vein isolation (PVI)
- Ablation that encircles the pulmonary veins so that their electrical activity cannot reach the left atrium. m39, m40
- Pulmonary vein reconnection
- Recovery of conduction between a pulmonary vein and the left atrium after isolation, usually through a gap in the lesion line. m53
- Quantization (integer)
- Storing a model’s weights or activations, or a signal’s samples, as integers with a scale and zero point instead of floating-point numbers; it saves memory and computation at the cost of rounding error. m47
- Quantization noise
- The rounding error of an analog-to-digital converter, with standard deviation one step over the square root of 12. m04
- Quantum noise
- Image noise caused by the random arrival of photons; with Poisson counts its standard deviation is the square root of the mean, so noise falls only as the square root of dose. m49
- Quasi-identifier
- A field that does not name a person on its own but can identify them in combination with others, such as age, sex, region or a procedure date. m47
- Quasi-static approximation
- Treating bioelectric fields as if they followed their sources instantly, neglecting capacitive, inductive and propagation effects in the volume conductor. m02
- Radial basis function
- A function of the distance to a centre, such as a Gaussian; sums of them represent smooth maps. m42, m18
- Radon transform
- The set of line integrals of an image over all lines in the plane, indexed by angle and offset; a CT scanner measures it. m51
- Randomized controlled trial (RCT)
- A study that assigns participants at random to an intervention or a comparison, so that differences in outcome can be attributed to the intervention. m45
- Receiver operating characteristic (ROC) curve
- Sensitivity plotted against one minus specificity as a decision threshold moves; used to choose cutoffs and compare classifiers. m23
- Receptive field
- The stretch of input that one output of a network depends on. m37
- Reciprocity
- Helmholtz’s result that the potential a source produces at an electrode equals the source dotted with the field the electrode would create by injecting unit current. m02
- Recovery coefficient
- The fraction of a structure’s true uptake (or contrast) that a blurred image measures; well below 1 for structures about as thin as the resolution. m54
- Recovery curve
- Conduction time (or another response) against the time since the tissue last conducted, measured with premature beats. m11
- Reentry
- A wave that keeps re-exciting tissue by travelling around a circuit, fixed or functional, back to where it started. m10, m22
- Reference catheter
- A catheter left in a stable position whose motion is used to correct other positions for shared motion. m19
- Reference electrogram
- A stable electrogram, often from the coronary sinus, whose annotated time on each beat is subtracted from local times to give local activation times. m22
- Reference level (radiation dose)
- A benchmark for the typical dose of a procedure type, set from surveys (for example the third quartile of recorded doses); used to compare and optimize practice, not as a limit for one patient. m49
- Reference standard
- The best available method for deciding whether a condition is present, against which a test or model is judged; its own errors limit any measured accuracy. m45
- Refractory period
- The time after an action potential during which a stimulus cannot start another one, or starts only a weaker one. m08
- Regularization
- Adding information or constraints, such as a penalty on solution size, so that an ill-posed problem gives a stable answer. m07
- Remote monitoring
- Sending data from an implanted or wearable device to a clinic, where alerts are reviewed. m38
- Repetition time (TR)
- The time between successive excitations in an MRI sequence; a real-time frame takes about the number of lines times TR. m53
- Respiratory gating
- Accepting positions or signals only within a window of the breathing cycle, usually near end expiration. m19
- Restitution
- The dependence of action potential duration (or conduction velocity) on the preceding diastolic interval. m08
- Retrieval-augmented generation (RAG)
- Answering with a language model after giving it passages retrieved from a chosen document collection, so that the answer can rest on, and cite, those passages. m43
- Rotor
- A spiral wave of excitation turning around a phase singularity. m10
- RR interval
- The time between two successive R peaks of the ECG, one ventricular beat to the next. m37
- Sampling density
- The number of mapped points per unit area, often given in points per cm²; it sets the typical distance to the nearest point. m26
- Scar
- Fibrous tissue that has replaced heart muscle. It conducts poorly or not at all. m03, m23
- Segmentation
- Labelling which voxels of an image belong to a structure, such as the left atrium. m40, m21
- Self-supervised learning
- Learning representations from unlabeled data with a training task built from the data themselves, such as matching two views of one recording. m43
- Sensor coil
- A small coil in a catheter whose induced voltage measures the component of an oscillating magnetic field along its axis. m16
- Single-fault condition
- Operation with one means of protection failed (for example an open protective earth), under which safety limits must still hold. m15
- Single-photon emission computed tomography (SPECT)
- Tomographic imaging of a tracer that emits single gamma photons, whose direction is set by a collimator. m54
- Singular value decomposition (SVD)
- Writing a matrix as a sum of rank-one terms s u vᵀ with orthonormal vectors and non-negative singular values s. m07
- Sinogram
- The Radon transform drawn as an image: one row of line integrals per projection angle. A point traces a sine curve in it. m51
- Sinus rhythm
- The normal heartbeat, started by the heart’s own pacemaker. m03
- Softmax
- The function that turns a vector of scores into probabilities that are positive and sum to one. m37
- Solid angle
- The area a surface covers on a unit sphere centered on the observer, in steradians; a full sphere is 4 pi. m01
- Source and sink
- Places where current leaves cells into the extracellular space (source) or enters them from it (sink); every source has a matching sink. m01
- Source–sink mismatch
- Too little depolarizing current from an excited region to bring a larger region of resting tissue to threshold, as when a wave leaves a narrow channel. m09
- Spatial frequency
- How quickly a pattern varies in space, in rad/mm. Fine detail has high spatial frequency. m03
- Spatial Nyquist limit
- The highest spatial wavenumber, pi divided by the electrode spacing, that a regular array of electrodes can represent without aliasing. m13
- Spatial resolution
- How close two structures can be and still appear separate in an image; often quoted as the FWHM of the point-spread function. m48
- Spectral leakage
- Power from one frequency spreading into others because a finite record is a windowed signal; window choice controls it. m05
- Speed-of-sound assumption
- Ultrasound systems convert echo time to depth with one assumed speed of sound, commonly 1540 m/s; tissue that is faster or slower is drawn shorter or longer. m50
- Spreading resistance
- The resistance between a small electrode and a distant return, concentrated in the medium near the electrode; it scales as one over the electrode’s size. m13
- Standardized uptake value (SUV)
- PET activity concentration divided by the injected activity per unit body weight, after decay correction; 1 means uniform distribution. m54
- Surface registration
- Registration that fits one surface to another without paired points, for example by ICP; some motions slide a surface along itself without changing the fit. m21
- Surrogate model
- A fast approximation of a slow model, fitted to runs of it and valid only in the range it was trained on. m42
- Surrogate signal
- A measurable signal, such as diaphragm or reference-catheter motion, used to estimate motion that cannot be measured directly. m19
- T1 (longitudinal relaxation time)
- The time constant with which magnetization along the field recovers after it is disturbed; gadolinium shortens it. m52
- Target registration error (TRE)
- The error after registration at a point not used for the fit, such as an ablation target; it grows with distance from the landmarks. m55, m21
- Taubin smoothing
- Mesh smoothing that alternates a shrinking and an inflating Laplacian step, so noise is removed without the overall shrinkage of plain Laplacian smoothing. m20
- Temporal resolution
- How short a time span one image represents. In cardiac CT it is about half the gantry rotation time with one X-ray source and a quarter with two. m51
- Test-retest variability
- How much a measurement changes when repeated on the same subject with nothing changed; it sets the smallest change that can be trusted. m54
- Tetralogy of Fallot
- A congenital heart defect affecting the right ventricular outflow and the ventricular septum. After surgical repair, ventricular tachycardia can use isthmuses bounded by patches and scars. m51
- TF-IDF
- Term frequency times inverse document frequency: a word weight that grows with how often the word appears in a document and with how rare it is across the collection. m43
- Thermal noise
- Voltage noise of a resistance R at temperature T, with density sqrt(4kTR) per root hertz (Johnson-Nyquist noise). m14
- Tikhonov regularization
- Least squares plus a penalty λ² times the squared size of the solution (or of its derivatives). m07, m32
- Token
- A unit of text, often a word or part of a word, that a language model reads and writes. m43
- Tool schema
- The declared arguments of an agent’s tool, with their types, units and allowed ranges, checked before the tool runs. m44
- Transesophageal echocardiography (TEE)
- Ultrasound imaging of the heart from a probe in the esophagus, behind the left atrium. m50
- Transformer
- A neural-network architecture built from attention layers; the basis of most large language models and of many ECG foundation models. m43
- Transmembrane voltage
- The voltage across a cell membrane, inside minus outside. m03
- Transseptal puncture
- Crossing from the right atrium to the left atrium through the interatrial septum, usually at the fossa ovalis, with a needle and sheath. m50
- Transthoracic echocardiography (TTE)
- Ultrasound imaging of the heart from a probe on the chest wall. m50
- Triangle mesh
- A surface stored as vertices, edges and triangular faces; mapping systems store the chamber geometry this way. m06
- Triangulation (conduction velocity)
- Estimating the activation-time gradient from three points by solving two linear equations; thin triangles amplify timing error. m25
- Truncated SVD
- Regularization that keeps only the components of the largest singular values. m07
- Unipolar electrogram
- The signal of one electrode measured against a distant reference electrode. It sees local and far-field activity. m03
- Velocity vector
- Conduction speed with direction: the activation-time gradient divided by its squared magnitude. m25
- Ventricle
- One of the two lower chambers of the heart. m03
- Ventricular tachycardia
- A fast rhythm that starts in the ventricles. m03, m41
- Voltage map
- A map colored by the peak-to-peak electrogram amplitude at each site, unipolar or bipolar, often with cutoffs for low voltage. m23
- Volume conductor
- The conducting medium (tissue and blood) through which currents spread from their sources to the electrodes. m03, m01, m02
- Wavefront curvature
- How sharply an activation front bends; convex fronts travel more slowly than flat ones. m09
- Wavelength of excitation
- Conduction velocity times refractory period: the length of tissue a wave keeps refractory. m10
- Wavelet transform
- A time-frequency analysis that correlates the signal with scaled copies of a short wavelet, using short windows for high frequencies and long ones for low. m05
- Welch’s method
- Estimating a power spectrum by averaging the spectra of several segments, which lowers the variance at the cost of resolution. m05
- Wenckebach periodicity
- Progressive lengthening of conduction time over successive beats until one beat is blocked, then repetition. m11
- WFDB format
- The Waveform Database format used by PhysioNet: integer samples in a signal file and a text header that gives the sampling rate, gain, baseline and units needed to convert them to physical values. m47
- Wilson central terminal (WCT)
- The average of the three limb-electrode potentials, used as the reference for unipolar chest leads. m12, m14
- Window of interest
- The span of time around the reference within which local activations are annotated; in a reentry it selects which cycle’s activation each site shows. m22
- Zero-phase filtering
- Filtering a record forward and then backward so that phase shifts cancel; it removes filter delay but needs the whole record, so it is offline only. m04