Terms used in the course, in alphabetical order. Each entry names the modules that use it. Clinical terms are defined in plain words for engineers; they are not clinical guidance.

Ablation
Destroying a small area of heart tissue on purpose, for example by radiofrequency heating, to stop an arrhythmia. m03
Action potential
The rise and fall of the transmembrane voltage when a heart cell is excited. m03, m08
Action potential duration (APD)
How long a cell stays depolarized after an upstroke, measured at a stated repolarization level. m08
Activation map
A map colored by local activation time, interpolated between the mapped points. m24
Activation wavefront
The moving boundary between excited and resting tissue. Also called the front. m03
Active catheter tracking
Locating small receiver coils at a catheter’s tip from MR signals so the scanner can follow it and steer the imaging plane; reported at up to 50 updates per second. m53
Adversarial example
An input changed slightly, often invisibly, so that a model gives a different output; for ECG models, a small added waveform. m46
AH interval
Time from the atrial to the His deflection on the His-bundle recording; mostly conduction through the AV node. m11
AI agent
A program in which a model, usually a large language model, chooses step by step which tool to call, reads the result and continues until a goal is met or a step limit is reached. m44
Air kerma
The kinetic energy that X-rays release per unit mass of air, in gray (Gy); measured at a reference point it relates to the dose at the patient’s skin. m49
Aliasing
The folding of frequencies above half the sampling rate onto lower ones; once sampled, the two cannot be told apart. m04
Alternans
Beat-to-beat alternation of action potential duration (or another quantity) at a steady pacing rate. m10
Analytic signal
A complex signal whose real part is the original and whose imaginary part is its Hilbert transform; its angle gives an instantaneous phase, meaningful when the signal is narrowband. m05
Anisotropy
Direction dependence of a tissue property: heart muscle conducts faster, and has lower resistance, along its fibers than across them. m01, m09
Applied part
The part of medical electrical equipment that contacts the patient in normal use; IEC 60601-1 classes include CF (cardiac floating) for direct cardiac connection. m15
Arrhythmia
An abnormal heart rhythm. m03
Atrial fibrillation (AF)
A fast, irregular rhythm of the upper chambers, without organized atrial activity. m37, m10
Atrium (plural atria)
One of the two upper chambers of the heart. m03
Attention
In a neural network, a weighted average of value vectors whose weights come from how well a query matches each key, normalized by a softmax. The building block of the transformer. m43
Audit log
An append-only record of every step an agent took: the tool, its arguments, the policy decision and the result, so a person can check the run afterwards. m44
Augmented limb leads
aVR, aVL and aVF: each limb electrode measured against the mean of the other two. m12
Average precision (PR AUC)
The area under the precision–recall curve, averaged over thresholds. Unlike the ROC AUC it falls when positives are rare. m39
Average symmetric surface distance (ASSD)
The mean distance between two surfaces, measured from each to the other; a typical-error measure for segmentations. m40
Basket catheter
A catheter whose electrodes sit on flexible splines that open into a cage filling a chamber, for simultaneous recording over a wide area. m13
Bayesian optimization
Choosing where to measure next by modelling an unknown function with a probabilistic surrogate, such as a Gaussian process, and maximizing an acquisition function. m41
Bidomain model
A tissue model with separate intracellular and extracellular potentials and conductivities, coupled through the membrane. m09
Bipolar electrogram
The difference between the signals of two nearby electrodes. It acts like a spatial derivative: it keeps local detail and rejects distant activity, but depends on the bipole’s orientation to the wavefront. m03, m13, m23
C-arm
A fluoroscopy system with the X-ray source and detector on opposite ends of a C-shaped arm that rotates around the patient. m55
Cable equation
The equation for voltage along a passive fiber with leaky membrane and axial resistance. m09
Cardiac sarcoidosis
Inflammation of the heart by granulomas, which can cause conduction block and ventricular tachycardia; FDG PET can show active inflammation. m54
Cavotricuspid isthmus (CTI)
The strip of right atrial tissue between the inferior vena cava and the tricuspid valve; the usual ablation target for typical atrial flutter. m50
Central-slice theorem
The 1-D Fourier transform of a projection equals the 2-D Fourier transform of the image along the line through the origin at the same angle; it is why projections at enough angles determine a CT slice. m51
Clique
A small group of electrodes, such as a triangle or a square, whose bipoles are combined to estimate the local electric field. m03, m25
Cohen’s kappa
Agreement between two raters or methods beyond what chance would give: (observed − expected agreement) / (1 − expected agreement). m53
Collocation point
A point where a governing equation is checked during fitting; it needs no measurement. m42
Common-mode rejection ratio (CMRR)
The ratio of an amplifier’s gain for the difference between its inputs to its gain for their mean, often in decibels. m14
Common-mode voltage
The mean of the voltages at a differential amplifier’s two inputs; on the body it is mostly picked up from the mains. m14
Complex conductivity
Conductivity plus the displacement-current term, sigma + j omega epsilon0 epsilon_r; its phase grows with frequency and differs between tissues. m18
Computed tomography (CT)
Imaging that measures X-ray attenuation along many lines through the body and reconstructs cross-sectional slices from those measurements. m51
Concealed conduction
An impulse that is blocked but still resets the excitability of the tissue it entered, so it affects the next beat. m11
Concealed fusion
Entrainment of a tachycardia with no change in the surface QRS (or P wave) morphology during pacing. m11
Concordance index (C index)
For a risk score, the share of comparable patient pairs in which the patient with the earlier event had the higher score; 0.5 is chance. m40
Condition number
The ratio of a matrix’s largest to smallest singular value; it bounds how much a relative error in the data can grow in the solution. m07, m13
Conducting channel
A corridor of surviving tissue through scar that may carry a reentrant ventricular tachycardia; post-processing software can identify candidate channels on LGE. m52
Conduction velocity
The speed at which the activation wavefront travels through tissue, in m/s (1 m/s = 1 mm/ms). m03, m42, m06, m09, m25
Conformal prediction
A method that turns a model’s scores into a set of possible answers that contains the true one at a chosen rate, using a held-out calibration set. m46
Contact force
The force between a catheter tip and the tissue, measured from the deformation of a small spring element near the tip. m13, m23
Contrast-to-noise ratio (CNR)
The difference between two tissues’ mean signals divided by the noise standard deviation; it decides whether they can be told apart, and averaging n independent samples multiplies it by the square root of n. m48, m49
Contrastive learning
Self-supervised learning that pulls the embeddings of matched pairs, such as two recordings of one patient, together and pushes unmatched pairs apart. m43
Convex hull
The smallest convex set containing a set of points; it cannot follow concavities. m20
Core-conductor model
A model of a cell strand in which intracellular current flows along the strand and leaks out through the membrane; it gives membrane current as the second derivative of transmembrane voltage. m01
Cotangent Laplacian
The standard discretization of the Laplace–Beltrami operator on a triangle mesh: each edge is weighted by half the sum of the cotangents of the two angles opposite it. m06
Cross-entropy
The loss most classifiers minimize: minus the log of the probability given to the true class. m37
Current dipole
A current source and an equal sink very close together; its strength is current times separation, and its potential falls as one over distance squared. m01
CUSUM
Cumulative sum: a running sum of deviations above a target, reset at zero, that raises an alarm when it crosses a threshold. Used to detect a lasting shift in a monitored quantity. m45
Cycle length
The time between successive activations in a repeating rhythm, in ms. m22
Data augmentation
Training on altered copies of the data (for example, flipped or noisier signals) so a model ignores the alteration. m37
Dataset shift
Any difference between the data a model was trained on and the data where it is used: in class mix (prior shift), in the recordings (covariate shift) or in what labels mean. m37
De-identification
Removing or coarsening information in health data so that it no longer identifies a person, for example by removing direct identifiers and grouping quasi-identifiers; the rules differ by country. m47
Decapolar catheter
A catheter with ten electrodes along its shaft. m03, m39
Deceleration zone
A region of crowded isochrones on an activation map, defined in one study as more than 3 isochrones within a 1-cm radius. m24
Diastolic interval (DI)
The time from the end of one action potential to the start of the next. m08
Dice coefficient
Overlap of a predicted and a true region: twice the shared area over the sum of the two areas, from 0 (no overlap) to 1 (identical). m39, m40, m21
Dielectric imaging
Building anatomy and tissue information from how electrical fields are shaped by the conductivity and permittivity of what surrounds the electrodes. m18
Discrepancy principle
Choosing a regularization parameter so that the model fits the data only as closely as the noise level allows. m07, m32
Discrete Picard condition
The requirement that the data’s coefficients in the singular vectors fall faster than the singular values; without it no regularized solution is good. m07
Dominant frequency
The frequency of the largest spectral peak in a search band; for electrograms it tracks the activation rate only after suitable preprocessing. m05, m29
Double layer
A surface covered with current dipoles pointing normal to it; a uniform double layer’s potential is proportional to the solid angle it subtends. m01
Drift
Slow change of reported positions over a procedure, for example from changing chest impedance; estimated by revisiting landmarks. m19
Driven right leg
A circuit that senses the body’s common-mode voltage, inverts and amplifies it, and drives it back through the right-leg electrode to reduce it. m14
Duty cycle
The fraction of time a gate accepts data. m19
Early meets late
In an activation map of a reentry, the place where the earliest and latest times sit side by side; it marks where the window of interest wraps, not necessarily a block line. m22
ECG gating
Using the electrocardiogram to keep only imaging data from a chosen part of the cardiac cycle, so that the moving heart appears still. m51
Eddy currents
Currents induced in a conductor by a changing magnetic field; their own field distorts magnetic localization. m16
Edema (myocardial)
Excess water in heart tissue after injury such as ablation; bright on T2-weighted MRI, it includes reversible injury and fades over time. m53
Effective refractory period (ERP)
The shortest coupling interval at which a premature stimulus still captures the tissue. m08
Eikonal equation
The equation v |grad T| = 1 linking an activation-time map T to the local conduction speed v. m42, m06
Einthoven triangle
The idealization in which the three limb leads have lead vectors 60 degrees apart around a centric heart vector. m12
Electrical remodeling
Changes in ion channels and action potentials caused by an arrhythmia itself, such as shorter atrial refractoriness during atrial fibrillation. m08
Electrogram
The voltage an electrode records from currents crossing cell membranes nearby. m03
Embedding
A vector that represents an input, such as a beat or a sentence, so that similar inputs get nearby vectors. m43
Endocardium
The inner surface of the heart wall. m03
Ensemble
A set of models trained with different random choices; their spread is used as an uncertainty estimate. m42
Entrainment
Capture of a reentrant circuit by pacing slightly faster than the tachycardia, without stopping it. m11
Error budget
A list of the error sources in a chain of steps with their sizes; independent, unbiased errors add in quadrature, and biases add directly. m55, m26
Euler characteristic
V − E + F for a mesh with V vertices, E edges and F faces; 2 for a closed surface without handles, one less for each hole cut into it. m06
Excitability
The ability of a membrane to respond to a stimulus above threshold with a full action potential. m08
Excitable gap
The recovered tissue ahead of a reentrant front: circuit length minus wavelength, or cycle length minus refractory period. m10
Expected improvement
An acquisition function: the average amount by which a new measurement is expected to beat the best value so far. m41
External validation
Testing a frozen model on data from another place or time that played no part in training or tuning. m37
Extracellular volume (ECV)
The fraction of a tissue’s volume outside cells. MRI estimates it from T1 before and after gadolinium and the hematocrit; scar and fibrosis raise it. m52
Far-field signal
Activity that an electrode picks up from tissue that is not directly under it. m03, m01, m22
Fast marching method
An algorithm that solves the eikonal equation by advancing a front from the source in order of arrival, on grids or triangle meshes. m06
FDG (fluorodeoxyglucose)
A glucose analog labeled with fluorine-18; cells that use much glucose, including inflammatory cells, take it up and show on PET. m54
Feature attribution
A score for how much each input feature, such as each sample of an ECG, contributed to a model’s output for one input. m46
Federated learning
Training one model across sites that keep their data; only model updates are shared. m37
Fibrosis
Replacement of heart muscle by connective tissue. In the atria it can be estimated on late-gadolinium MRI. m40
Fiducial localization error (FLE)
The error in locating a landmark used for registration, in either the image or the map. m55, m21
Fiducial registration error (FRE)
The residual distance between registered landmarks after a fit; its expected value depends on the number of landmarks and FLE, not on their layout. m55, m21
Field generator
The set of coils of known position and orientation that a magnetic localization system drives to create its reference fields. m16
Field scaling
Calibrating an impedance-based localization field with electrodes of known spacing, so measured potential differences give true distances. m17
Fill distance
The largest distance from any point of a surface to its nearest sample point or electrode; a measure of coverage. m13
Fill threshold
The distance from the nearest mapped point beyond which a mapping system leaves the surface uncolored. m24
Filter factor
The weight a regularization method gives each singular component of the solution: 1 or 0 for truncated SVD, s²/(s² + λ²) for Tikhonov. m07
Filtered back-projection
CT reconstruction that filters each projection with a ramp in frequency and smears it back along its lines; without the filter the image is blurred. m51
FitzHugh–Nagumo model
A two-variable model of an excitable membrane, with a fast voltage-like variable and a slow recovery variable. m08
Fluoroscopy
Continuous or pulsed X-ray imaging that shows a moving shadow image, used to guide catheters, sheaths and leads. m49
Focal loss
A cross-entropy that down-weights examples the model already gets right, so training attends to rare and hard ones. m39
Focal source
A small region from which activation spreads outward. m03, m10
Forward problem
Computing body-surface (or electrode) potentials from known cardiac sources and a model of the torso. m12
Fossa ovalis
The thin central part of the wall between the two atria, where a needle usually crosses from the right atrium to the left. m50
Foundation model
A large model pretrained on broad, mostly unlabeled data and then adapted to many downstream tasks. m43
Frontal axis
The direction of the heart vector in the frontal plane, usually from leads I and aVF; 0 degrees toward the left arm, +90 toward the feet. m12
Full width at half maximum (FWHM)
The width of a blur between the two points where it falls to half its peak; for a Gaussian, about 2.355 standard deviations. Used to quote the spatial resolution of scanners. m48, m51, m54
Gadolinium contrast
An MRI contrast agent that shortens T1 in proportion to its concentration and stays outside cells, so it accumulates where there is more extracellular space. m52
Gating variable
A number between 0 and 1 for the open fraction of one kind of channel gate, opening and closing at voltage-dependent rates. m08
Gaussian mixture model
A probability model that describes data as a weighted sum of normal distributions; fitted with the EM algorithm to find clusters without labels. m39
Gaussian process
A probability model over functions; given measurements, it gives a mean estimate and an uncertainty at every other point. m41, m26
Generalized cross-validation (GCV)
Choosing a regularization parameter by minimizing a predicted-error function of the data, without a noise estimate. m07
Geodesic distance
The length of the shortest path between two points that stays on a surface. m41, m06
Graph convolution
A network layer that mixes each point’s features with those of its neighbors on a graph, such as the vertices of a mapping mesh. m39
Grating lobe
A false, full-strength copy of an array’s beam that appears when its elements are spaced more than half a wavelength apart for the steering used; echoes from it are drawn in the wrong direction. m50
Group delay
How long a filter delays the envelope of a signal at a given frequency; causal filters shift features such as activation times by it. m04
Guardrail
A rule enforced outside the model that limits what an agent may do, such as a tool allowlist, a schema check or a human approval step. m44
Half-life
The time for a radioactive tracer’s activity to halve; 109.73 minutes for fluorine-18. m54
Half-value layer (HVL)
The thickness of a material that halves the intensity of an X-ray beam; for a narrow beam of one energy it is ln 2 divided by the attenuation coefficient. m49
Hallucination
Output of a language model that reads as fluent and confident but is false or unsupported, such as an invented number or reference. m43
Harmonic interpolation
Filling in values on a mesh by solving Laplace’s equation between the known vertices; it cannot overshoot the data. m24
Hausdorff distance
The largest distance from any point of one surface to the other surface, in either direction; one stray island sets it. m40, m20
Heart vector
The single equivalent dipole that sums the heart’s current sources at one instant. m12
Heat method
A way to compute geodesic distance on a mesh: diffuse heat briefly from the source, normalize its gradient, then solve a Poisson equation. m06
Hematocrit
The fraction of blood volume made up of red cells; gadolinium in blood is confined to the rest, the plasma. m52
High-density grid
A catheter whose electrodes form a planar array, so bipoles can be formed along two perpendicular directions. m13
His bundle
The bundle of conducting fibers that carries activation from the AV node to the bundle branches and Purkinje system. m11
Hopf bifurcation
The point at which a steady state loses stability and small oscillations appear; in excitable cells, the onset of repetitive firing. m08
Hounsfield unit (HU)
The CT number scale, 1000 (μ − μ_water)/(μ_water − μ_air): water is 0 and air −1000. Values depend on the X-ray beam’s energy. m51
Human approval step
A point in an agent’s run where a person must accept or decline an action, such as any write, before it is carried out. m44
HV interval
Time from the His deflection to the earliest ventricular activation; conduction through the His–Purkinje system. m11
Hybrid localization
Correcting impedance-based electrode positions with points where a magnetic sensor gives the true position, so every electrode can be shown accurately. m18
Ill-posed problem
A problem whose solution fails to exist, to be unique or to depend continuously on the data; small data errors then cause large solution errors. m07
Image integration
Bringing a CT or MRI anatomy into a mapping system’s coordinates by segmentation and registration, so the catheter can be shown in it. m21
Image intensity ratio (IIR)
An LGE wall intensity divided by the mean blood-pool intensity, used to compare scans that have no absolute intensity scale; thresholds on it define atrial fibrosis or scar. m52
Impedance localization
Locating catheter electrodes from the potentials that high-frequency currents between skin patches make at them. m17
Implantable loop recorder (ILR)
A small device placed under the skin that records the ECG for years and sends alerts when its detector fires. m38
Instrumentation amplifier
A differential amplifier with a high-impedance, gain-setting input stage ahead of a difference stage, which multiplies its common-mode rejection. m14
Integrated gradients
A feature-attribution method that averages the model’s gradient along a straight path from a baseline input to the actual input; the attributions add up to the change in output. m46
Interventional MRI
Performing a procedure, such as catheter ablation, inside an MRI scanner with real-time images instead of fluoroscopy. m53
Intracardiac echocardiography (ICE)
Ultrasound imaging from a catheter inside the heart, used during procedures to see anatomy and build shells. m40
Intrinsic deflection
The steepest downstroke of a unipolar electrogram, used to mark local activation. m03, m22
Inverse problem
Recovering sources from measurements, the reverse of the forward problem that predicts measurements from sources. m07, m32
Inverse-distance weighting (IDW)
Interpolation by a weighted mean of the data with weights proportional to distance to the power -p; for p > 1 it is flat at each data point. m24
Inverse-variance weighting
Combining independent unbiased estimates with weights proportional to one over their variances; the result is at least as precise as the best one. m18
Inversion recovery
An MRI preparation that flips the longitudinal magnetization with a 180° pulse and reads the image while it recovers; a tissue has no signal at T1 ln 2. m52
Iodinated contrast
An iodine-containing agent injected into the blood; iodine attenuates X-rays strongly, so the blood pool stands out from the heart wall on CT. m51
Isocenter
The point a C-arm rotates about; structures there are imaged at a fixed magnification. m55
Isochrone
A line joining sites activated at the same time; isochrones a fixed time step apart are closer where conduction is slower. m24
Isolation barrier
The insulation and spacing that separate patient-connected circuits from mains-powered circuits; its capacitance still passes a small leakage current. m15
Isthmus (VT)
A narrow channel of conducting tissue, bounded by scar or anatomical barriers, that a reentrant ventricular tachycardia passes through. m51
Iterative closest point (ICP)
Surface registration that alternately pairs each point with its nearest point on the other surface and fits a rigid transform to those pairs; it converges to the nearest local minimum. m21
Jaccard index
Overlap of two regions as intersection over union; Dice = 2J/(1 + J). m40, m21
k-anonymity
A property of a released table: every record shares its quasi-identifier values with at least k − 1 others, so a match by those values points to at least k records. m47
Kerma-area product (KAP)
Air kerma times the area of the X-ray beam, in Gy·cm²; the same at any distance along the beam, so it describes the total radiation delivered. Also called dose-area product. m49
L-curve
A log-log plot of solution size against residual size over all regularization parameters; its corner is one way to choose the parameter. m07, m32
Landmark registration
Registration from paired points located in both coordinate systems, solved in closed form by least squares with an SVD. m21
Laplace–Beltrami operator
The Laplacian on a curved surface: the divergence of the surface gradient. m06
Large language model (LLM)
A transformer trained on very large amounts of text to predict the next token; it can follow instructions and answer questions in text. m43
Larmor frequency
The frequency at which nuclear magnetization precesses in a magnetic field: 42.58 MHz per tesla for protons, about 64 MHz at 1.5 T. m52
Late gadolinium enhancement (LGE)
MRI taken some minutes after a gadolinium contrast injection, when scar and fibrosis appear bright. m40, m52
Lead field
An electrode’s sensitivity to sources everywhere: by reciprocity, the gradient of the potential it would create by injecting unit current. m02, m12
Lead vector
The vector L for which a lead’s voltage equals L . p for a dipole p; it depends on the electrodes and the torso. m12
Leakage current
Current that flows from equipment through an unintended path, such as the patient to ground, in normal use or with a single fault. m15
Leave-one-out cross-validation
Scoring a model by predicting each data point from all the others; for a Gaussian process it needs only one matrix inverse. m26
Leave-one-patient-out
A test scheme that trains on all patients but one and tests on that one, in turn, so no patient is in both sets. m39
Left atrial appendage occlusion (LAAO)
A catheter procedure that closes the left atrial appendage with an implanted device. m50
Length constant
The distance over which a held voltage decays to 1/e along a passive cable, the square root of membrane over axial resistance. m09
Length scale (kernel)
The distance over which a Gaussian process’s covariance falls off, which sets how far each point’s influence reaches. m26
Lesion gap
A segment of an intended ablation line where the tissue was not destroyed, through which conduction can recover. m53
Lesion metabolic activity
A PET measure of a lesion: its partial-volume-corrected mean uptake times its volume. m54
Levenberg–Marquardt method
An iterative method for nonlinear least squares that blends Gauss–Newton steps with damped gradient steps. m42, m16
Likelihood ratio
How much a test result changes the odds of disease: for a positive result, sensitivity divided by one minus specificity; for a negative result, one minus sensitivity divided by specificity. Post-test odds are pre-test odds times the likelihood ratio. m48
Line of block
A line across which activation does not pass directly, so sites on either side activate at very different times. m24
Linear probe
A linear classifier trained on the frozen features of a pretrained model, used to measure how much those features already encode. m43
Local activation time
The moment the activation wavefront reaches a given site. m03, m22
Local plane fit
Estimating the activation-time gradient by fitting a plane to the activation times of nearby points by least squares; more points reduce noise, a larger radius adds bias where activation curves. m25
Locally Delaunay edge
A mesh edge whose two opposite angles add up to at most 180°; its cotangent weight is then not negative. m06
Low-voltage area
A region whose bipolar voltage falls below a chosen cut-off. It is often read as scar, but the voltage also depends on the catheter and the wavefront. m03, m23
Macro-reentry
Reentry around a large circuit set by anatomy or scar, such as typical atrial flutter. m10
Macroshock
Electric shock from current entering and leaving through the skin, so it spreads through the body before reaching the heart. m15
Magnetic dipole
The far-field model of a small current loop, with moment m; its field falls as the cube of distance. m16
Magnetic resonance imaging (MRI)
Imaging of hydrogen nuclei in a strong magnetic field, with contrast from how their magnetization relaxes and position encoded by field gradients. m52
Magnetohydrodynamic voltage
A voltage induced when conducting blood flows across a strong magnetic field; it distorts the ECG and electrograms inside an MRI scanner. m53
Magnification (projection)
How much larger a structure appears on the detector than its true size: source-to-detector over source-to-object distance. m55
Marching cubes
An algorithm that extracts a triangle surface at a chosen level from volumetric data. m20
Membrane current
Current crossing the cell membrane, positive outward; ahead of a wavefront it flows out (a source) and behind it flows in (a sink). m01
Method of images
Solving a problem with a flat boundary by adding mirror sources that make the boundary conditions hold; for two conductors the mirror source has strength (sigma1 - sigma2)/(sigma1 + sigma2) times the original. m02
Microshock
Electric shock from current delivered directly to the heart, for example through a catheter, where microamperes can capture or fibrillate the ventricles. m15
Mitchell–Schaeffer model
A two-variable cardiac membrane model with one gated inward current and one outward current; its action potential duration has a closed form. m08
Model card
A short document released with a trained model that states its intended use, out-of-scope uses, training and evaluation data, metrics overall and by group, and limitations. m47
Monodomain model
A tissue model with one potential and one effective conductivity; exact in 1D or when both domains are equally anisotropic. m09
Morphological closing
Dilation followed by erosion with the same shape; it fills gaps and concavities narrower than the shape while keeping larger features. m20
Myocardial infarction
Death of an area of heart muscle caused by loss of its blood supply. m03
Negative predictive value (NPV)
The share of negative results that are truly negative; like the positive predictive value, it depends on prevalence. m48, m35
Nernst potential
The voltage at which an ion’s diffusion and electrical drift balance: (RT/zF) ln of the outside-to-inside concentration ratio. m08
Net benefit
In decision-curve analysis, true positives minus false positives weighted by the odds of the threshold probability, per patient. It compares a model with acting on everyone or no one. m45
Noise factor
Total input-referred noise power divided by the noise power of the source resistance alone; it is lowest when the source resistance equals the amplifier’s voltage noise over its current noise. m14
Noise floor
The smallest amplitude a measurement can show: with no signal, the peak-to-peak of noise in the measurement window. m23
Noncontact mapping
Reconstructing endocardial potentials from a multielectrode probe that floats in the cavity, by solving an inverse problem. m07
Normalized surface Dice (NSD)
The share of two surfaces lying within a chosen tolerance of each other. m40
Notch filter
A narrow band-stop filter used to remove mains interference at 50 or 60 Hz. m04
Nyquist frequency
Half the sampling rate: the highest frequency a sampled signal can represent without aliasing. m04
Omnipolar electrogram
A signal built from a clique by estimating the local electric field and reading it along the direction of propagation, which removes most orientation dependence. m03, m23, m25
Otsu’s threshold
The intensity cut that splits a histogram into two classes with the largest between-class variance; used as a data-driven threshold for scar. m52, m21
Out-of-distribution (OOD) input
An input unlike the data a model was trained on, such as a recording from a new device or with heavy noise, for which its outputs are not reliable. m46
Outflow tract
The part of a ventricle leading to its great artery (pulmonary artery on the right, aorta on the left); a common origin of idiopathic premature ventricular beats. m41
Oversampling
Sampling faster than needed and averaging; for white noise each doubling of the averaged samples gains about 3 dB, half a bit. m14
Pace mapping
Pacing from a catheter at candidate sites and comparing the paced 12-lead ECG with the ECG of the arrhythmia, to find where the arrhythmia starts. m41
Pacing
Stimulating the heart electrically from an electrode at a chosen site. m03
Partial-volume effect
A voxel, or a region blurred by the scanner, that holds more than one tissue reads a mixture of their values, so thin or small structures lose contrast and change apparent size. m51, m54, m48
Passive catheter visualization
Seeing a catheter on real-time MR images through the signal void or artifact its materials cause; it shows one plane at a time. m53
Patch pair
Two skin electrodes that drive a localization current through the chest; three roughly orthogonal pairs give three coordinates. m17
Peak-to-peak amplitude
The difference between the largest and smallest values of a signal in a window; the usual electrogram voltage. m23
Phase singularity
A point around which the phase of activation takes every value; a spiral wave turns around one. m10
Phase-sensitive inversion recovery (PSIR)
An LGE reconstruction that keeps the sign of the recovering magnetization, so it is much less sensitive to the chosen inversion time than a magnitude image. m52
Phased array
A row of transducer elements whose firing delays steer and focus the beam electronically, without moving the probe. m50
Photoplethysmography (PPG)
Measuring blood-volume changes in tissue with light; the basis of pulse sensing in watches and trackers. m38
Physics-informed neural network (PINN)
A neural network trained on a loss that adds the residual of a governing equation to the data misfit. m42
Point-spread function (PSF)
The image an imaging system makes of a single point: the blur it applies to everything it shows. m48
Poisson’s equation
The equation the potential obeys in a volume conductor, div(sigma grad phi) = -I_v: Ohm’s law plus current conservation. m02
Pose
Position together with orientation; a single sensor coil gives five of its six degrees of freedom. m16
Positive predictive value (PPV)
The share of positive results (alerts, flags) that are truly positive; it depends on prevalence. m37, m38, m48
Positron emission tomography (PET)
Imaging of a tracer that emits positrons; each annihilation sends two photons in opposite directions, detected together to define a line through the source. m54
Post-pacing interval (PPI)
Time from the last pacing stimulus to the next local activation at the pacing site after entrainment. m11
Pre-test probability
The probability of a condition before a test result is known, for example its prevalence in the group tested. m48
Predetermined change control plan (PCCP)
A plan, authorized with an AI-enabled device’s marketing submission, that describes planned changes to the device and how each will be validated, so those changes need no new submission. m45
Premature beat
A beat that comes earlier than expected, usually followed by a longer interval; a common cause of false AF alerts. m38
Premature ventricular contraction (PVC)
An extra beat that starts in a ventricle earlier than the next normal beat. m41
Prompt injection
Text inside a document or tool output that tries to make a language model follow new instructions, such as a hidden request to delete files. m44
Pulmonary vein isolation (PVI)
Ablation that encircles the pulmonary veins so that their electrical activity cannot reach the left atrium. m39, m40
Pulmonary vein reconnection
Recovery of conduction between a pulmonary vein and the left atrium after isolation, usually through a gap in the lesion line. m53
Quantization (integer)
Storing a model’s weights or activations, or a signal’s samples, as integers with a scale and zero point instead of floating-point numbers; it saves memory and computation at the cost of rounding error. m47
Quantization noise
The rounding error of an analog-to-digital converter, with standard deviation one step over the square root of 12. m04
Quantum noise
Image noise caused by the random arrival of photons; with Poisson counts its standard deviation is the square root of the mean, so noise falls only as the square root of dose. m49
Quasi-identifier
A field that does not name a person on its own but can identify them in combination with others, such as age, sex, region or a procedure date. m47
Quasi-static approximation
Treating bioelectric fields as if they followed their sources instantly, neglecting capacitive, inductive and propagation effects in the volume conductor. m02
Radial basis function
A function of the distance to a centre, such as a Gaussian; sums of them represent smooth maps. m42, m18
Radon transform
The set of line integrals of an image over all lines in the plane, indexed by angle and offset; a CT scanner measures it. m51
Randomized controlled trial (RCT)
A study that assigns participants at random to an intervention or a comparison, so that differences in outcome can be attributed to the intervention. m45
Receiver operating characteristic (ROC) curve
Sensitivity plotted against one minus specificity as a decision threshold moves; used to choose cutoffs and compare classifiers. m23
Receptive field
The stretch of input that one output of a network depends on. m37
Reciprocity
Helmholtz’s result that the potential a source produces at an electrode equals the source dotted with the field the electrode would create by injecting unit current. m02
Recovery coefficient
The fraction of a structure’s true uptake (or contrast) that a blurred image measures; well below 1 for structures about as thin as the resolution. m54
Recovery curve
Conduction time (or another response) against the time since the tissue last conducted, measured with premature beats. m11
Reentry
A wave that keeps re-exciting tissue by travelling around a circuit, fixed or functional, back to where it started. m10, m22
Reference catheter
A catheter left in a stable position whose motion is used to correct other positions for shared motion. m19
Reference electrogram
A stable electrogram, often from the coronary sinus, whose annotated time on each beat is subtracted from local times to give local activation times. m22
Reference level (radiation dose)
A benchmark for the typical dose of a procedure type, set from surveys (for example the third quartile of recorded doses); used to compare and optimize practice, not as a limit for one patient. m49
Reference standard
The best available method for deciding whether a condition is present, against which a test or model is judged; its own errors limit any measured accuracy. m45
Refractory period
The time after an action potential during which a stimulus cannot start another one, or starts only a weaker one. m08
Regularization
Adding information or constraints, such as a penalty on solution size, so that an ill-posed problem gives a stable answer. m07
Remote monitoring
Sending data from an implanted or wearable device to a clinic, where alerts are reviewed. m38
Repetition time (TR)
The time between successive excitations in an MRI sequence; a real-time frame takes about the number of lines times TR. m53
Respiratory gating
Accepting positions or signals only within a window of the breathing cycle, usually near end expiration. m19
Restitution
The dependence of action potential duration (or conduction velocity) on the preceding diastolic interval. m08
Retrieval-augmented generation (RAG)
Answering with a language model after giving it passages retrieved from a chosen document collection, so that the answer can rest on, and cite, those passages. m43
Rotor
A spiral wave of excitation turning around a phase singularity. m10
RR interval
The time between two successive R peaks of the ECG, one ventricular beat to the next. m37
Sampling density
The number of mapped points per unit area, often given in points per cm²; it sets the typical distance to the nearest point. m26
Scar
Fibrous tissue that has replaced heart muscle. It conducts poorly or not at all. m03, m23
Segmentation
Labelling which voxels of an image belong to a structure, such as the left atrium. m40, m21
Self-supervised learning
Learning representations from unlabeled data with a training task built from the data themselves, such as matching two views of one recording. m43
Sensor coil
A small coil in a catheter whose induced voltage measures the component of an oscillating magnetic field along its axis. m16
Single-fault condition
Operation with one means of protection failed (for example an open protective earth), under which safety limits must still hold. m15
Single-photon emission computed tomography (SPECT)
Tomographic imaging of a tracer that emits single gamma photons, whose direction is set by a collimator. m54
Singular value decomposition (SVD)
Writing a matrix as a sum of rank-one terms s u vᵀ with orthonormal vectors and non-negative singular values s. m07
Sinogram
The Radon transform drawn as an image: one row of line integrals per projection angle. A point traces a sine curve in it. m51
Sinus rhythm
The normal heartbeat, started by the heart’s own pacemaker. m03
Softmax
The function that turns a vector of scores into probabilities that are positive and sum to one. m37
Solid angle
The area a surface covers on a unit sphere centered on the observer, in steradians; a full sphere is 4 pi. m01
Source and sink
Places where current leaves cells into the extracellular space (source) or enters them from it (sink); every source has a matching sink. m01
Source–sink mismatch
Too little depolarizing current from an excited region to bring a larger region of resting tissue to threshold, as when a wave leaves a narrow channel. m09
Spatial frequency
How quickly a pattern varies in space, in rad/mm. Fine detail has high spatial frequency. m03
Spatial Nyquist limit
The highest spatial wavenumber, pi divided by the electrode spacing, that a regular array of electrodes can represent without aliasing. m13
Spatial resolution
How close two structures can be and still appear separate in an image; often quoted as the FWHM of the point-spread function. m48
Spectral leakage
Power from one frequency spreading into others because a finite record is a windowed signal; window choice controls it. m05
Speed-of-sound assumption
Ultrasound systems convert echo time to depth with one assumed speed of sound, commonly 1540 m/s; tissue that is faster or slower is drawn shorter or longer. m50
Spreading resistance
The resistance between a small electrode and a distant return, concentrated in the medium near the electrode; it scales as one over the electrode’s size. m13
Standardized uptake value (SUV)
PET activity concentration divided by the injected activity per unit body weight, after decay correction; 1 means uniform distribution. m54
Surface registration
Registration that fits one surface to another without paired points, for example by ICP; some motions slide a surface along itself without changing the fit. m21
Surrogate model
A fast approximation of a slow model, fitted to runs of it and valid only in the range it was trained on. m42
Surrogate signal
A measurable signal, such as diaphragm or reference-catheter motion, used to estimate motion that cannot be measured directly. m19
T1 (longitudinal relaxation time)
The time constant with which magnetization along the field recovers after it is disturbed; gadolinium shortens it. m52
Target registration error (TRE)
The error after registration at a point not used for the fit, such as an ablation target; it grows with distance from the landmarks. m55, m21
Taubin smoothing
Mesh smoothing that alternates a shrinking and an inflating Laplacian step, so noise is removed without the overall shrinkage of plain Laplacian smoothing. m20
Temporal resolution
How short a time span one image represents. In cardiac CT it is about half the gantry rotation time with one X-ray source and a quarter with two. m51
Test-retest variability
How much a measurement changes when repeated on the same subject with nothing changed; it sets the smallest change that can be trusted. m54
Tetralogy of Fallot
A congenital heart defect affecting the right ventricular outflow and the ventricular septum. After surgical repair, ventricular tachycardia can use isthmuses bounded by patches and scars. m51
TF-IDF
Term frequency times inverse document frequency: a word weight that grows with how often the word appears in a document and with how rare it is across the collection. m43
Thermal noise
Voltage noise of a resistance R at temperature T, with density sqrt(4kTR) per root hertz (Johnson-Nyquist noise). m14
Tikhonov regularization
Least squares plus a penalty λ² times the squared size of the solution (or of its derivatives). m07, m32
Token
A unit of text, often a word or part of a word, that a language model reads and writes. m43
Tool schema
The declared arguments of an agent’s tool, with their types, units and allowed ranges, checked before the tool runs. m44
Transesophageal echocardiography (TEE)
Ultrasound imaging of the heart from a probe in the esophagus, behind the left atrium. m50
Transformer
A neural-network architecture built from attention layers; the basis of most large language models and of many ECG foundation models. m43
Transmembrane voltage
The voltage across a cell membrane, inside minus outside. m03
Transseptal puncture
Crossing from the right atrium to the left atrium through the interatrial septum, usually at the fossa ovalis, with a needle and sheath. m50
Transthoracic echocardiography (TTE)
Ultrasound imaging of the heart from a probe on the chest wall. m50
Triangle mesh
A surface stored as vertices, edges and triangular faces; mapping systems store the chamber geometry this way. m06
Triangulation (conduction velocity)
Estimating the activation-time gradient from three points by solving two linear equations; thin triangles amplify timing error. m25
Truncated SVD
Regularization that keeps only the components of the largest singular values. m07
Unipolar electrogram
The signal of one electrode measured against a distant reference electrode. It sees local and far-field activity. m03
Velocity vector
Conduction speed with direction: the activation-time gradient divided by its squared magnitude. m25
Ventricle
One of the two lower chambers of the heart. m03
Ventricular tachycardia
A fast rhythm that starts in the ventricles. m03, m41
Voltage map
A map colored by the peak-to-peak electrogram amplitude at each site, unipolar or bipolar, often with cutoffs for low voltage. m23
Volume conductor
The conducting medium (tissue and blood) through which currents spread from their sources to the electrodes. m03, m01, m02
Wavefront curvature
How sharply an activation front bends; convex fronts travel more slowly than flat ones. m09
Wavelength of excitation
Conduction velocity times refractory period: the length of tissue a wave keeps refractory. m10
Wavelet transform
A time-frequency analysis that correlates the signal with scaled copies of a short wavelet, using short windows for high frequencies and long ones for low. m05
Welch’s method
Estimating a power spectrum by averaging the spectra of several segments, which lowers the variance at the cost of resolution. m05
Wenckebach periodicity
Progressive lengthening of conduction time over successive beats until one beat is blocked, then repetition. m11
WFDB format
The Waveform Database format used by PhysioNet: integer samples in a signal file and a text header that gives the sampling rate, gain, baseline and units needed to convert them to physical values. m47
Wilson central terminal (WCT)
The average of the three limb-electrode potentials, used as the reference for unipolar chest leads. m12, m14
Window of interest
The span of time around the reference within which local activations are annotated; in a reentry it selects which cycle’s activation each site shows. m22
Zero-phase filtering
Filtering a record forward and then backward so that phase shifts cancel; it removes filter delay but needs the whole record, so it is offline only. m04