Skip to main content

m29 · Complex maps: fractionation, phase mapping, rotor and driver detection

Warning

Education only. Not for diagnosis, treatment or any clinical decision.

Note

Draft, stage 2 of 6. The derivations are written, and each equation is checked by tests in tests/test_m29_complex_maps.py. Demos, code lab, questions and the remaining sections follow in later stages (plan: docs/outlines/m29-outline.md). Nothing here has had expert review yet. Whether rotors or other drivers sustain human atrial fibrillation, and whether ablating them helps, is contested; this module teaches how the maps are computed and tested, and gives no ablation advice.

In 60 seconds

During atrial fibrillation (AF) the atrium does not activate in one clean wave, so a single activation time per site stops making sense. “Complex maps” replace it with other summaries: how fragmented the local signal is, how fast it repeats, or where each site is in its cycle (its phase). Points around which phase winds through a full cycle are phase singularities, the signature of a rotor in theory. Every one of these maps depends on processing choices and on how far apart the electrodes are, so each must be tested on data where the truth is known before its output is trusted.

From the field

Planned in the outline (docs/outlines/m29-outline.md); written in a later stage.

Derivations

Symbols used throughout. \(x(t)\) an electrogram, sampled at \(f_s\) (Hz). \(X(f)\) its Fourier transform. \(t_k\) times of detected deflections. \(\varphi\) phase (rad). \(\mathcal{H}\) the Hilbert transform. \(v\) conduction velocity and \(\mathrm{CL}\) cycle length. \(\lambda_s\) spatial wavelength. \(n\) topological charge.

Fractionation

A complex fractionated atrial electrogram (CFAE) has many deflections, or continuous activity, in place of one clean deflection per beat. Such signals were proposed as ablation targets in a large series that mapped them during AF (Nademanee et al. 2004). To compare sites, a mapping system turns “fragmented” into a number. A simple one is the mean interval between deflections detected in a window:

\[ \overline{\Delta t} = \frac{t_N - t_1}{N - 1} \tag{1}\]

for \(N\) deflections at times \(t_1 < \dots < t_N\). Shorter means more fragmented. The number depends entirely on what counts as a deflection: the filter, the detection threshold, the minimum spacing between detections and the electrode spacing. Halving the threshold can double the count of deflections in the same signal (a test checks this on a synthetic train with large and small deflections). Commercial fractionation algorithms are not reproduced here.

Key takeaways

  • Fractionation indices count detected deflections; they inherit every choice made by the detector.
  • Report the detector settings with any fractionation map.

Dominant frequency and organization

If activation at a site repeats roughly every \(\mathrm{CL}\), its spectrum has a peak at \(1/\mathrm{CL}\) and at harmonics. The dominant frequency (DF) is the frequency of the largest peak within a search band \([f_{\text{lo}}, f_{\text{hi}}]\):

\[ \mathrm{DF} = \arg\max_{f \in [f_{\text{lo}},\, f_{\text{hi}}]} \lvert X(f)\rvert^2 \tag{2}\]

Raw bipolar electrograms are sharp, so most of their power sits at frequencies far above the activation rate, and preprocessing is a necessary step for them (Ng et al. 2007). A common chain is band-pass filter, rectify, then low-pass filter, so that each activation becomes one smooth bump. The tests show the effect: on a regular synthetic train, DF of the preprocessed signal equals \(1/\mathrm{CL}\). The origin of this chain and the filter values used in published work: TODO(source).

DF estimates depend on unipolar versus bipolar recording, bipolar spacing, preprocessing, far-field ventricular signals and window length; in one technical study, records longer than 2 s were needed for reliable DF (Ng et al. 2007). DF tracks the activation rate only when its assumptions hold; other features of the signal can move the peak (Ng and Goldberger 2007). DF maps built from electroanatomic points have been used to locate sites of high-frequency activity during AF (Sanders et al. 2005).

How regular is the signal? Everett and colleagues defined an organization index (OI) as the area under the first four harmonic peaks divided by the total area of the spectrum (Everett et al. 2001):

\[ \mathrm{OI} = \frac{\sum_{h=1}^{4} \text{area under the peak at } h \cdot \mathrm{DF}}{\text{total spectral area}} \tag{3}\]

A perfectly periodic signal puts all its power in harmonics, so OI is near 1; irregular timing spreads power and lowers OI. The width of the band counted as “under the peak” is a choice that must be reported.

Key takeaways

  • DF is the largest spectral peak after preprocessing; without preprocessing it can sit at a harmonic.
  • DF depends on recording type, spacing, far-field signals and window length.
  • OI measures how much power is in the harmonics of DF.

Phase

Phase places each sample on a cycle: \(0\) to \(2\pi\) per activation. The analytic signal adds the Hilbert transform of \(x\) as an imaginary part:

\[ z(t) = x(t) + i\,\mathcal{H}[x](t), \qquad \varphi(t) = \arg z(t) \tag{4}\]

For \(x = \cos\omega t\), \(\mathcal{H}[x] = \sin\omega t\), so \(z = e^{i\omega t}\) and the phase advances evenly, as the tests check. Optical recordings of transmembrane voltage give smooth, nearly periodic signals, and phase maps built from them showed phase singularities during fibrillation (Gray et al. 1998).

Contact electrograms are different. For unipolar atrial electrograms, the phase from the Hilbert transform depends on whether a deflection is one- or two-phased, on the ratio of its R and S waves, and on noise (Kuklik et al. 2015). Kuklik and colleagues proposed rebuilding the electrogram first from sine wavelets whose amplitudes are proportional to the negative slope of the signal (Kuklik et al. 2015). With \(w(\tau) = \max\!\left(0, -\dot x(\tau)\right)\) and a wavelet \(\psi\) that is one period of a sine of length \(T\):

\[ s(t) = \int w(\tau)\, \psi(t - \tau)\, d\tau, \qquad \psi(u) = \sin\!\left(\frac{2\pi u}{T}\right) \text{ for } \lvert u\rvert \le \tfrac{T}{2}, \ 0 \text{ otherwise} \tag{5}\]

then take the Hilbert phase of \(s\). In this course \(T\) is set to the cycle length. Each activation, whatever its shape, then adds one phase cycle (a test checks this on signals with random R/S ratios and noise).

Key takeaways

  • Hilbert phase is clean for smooth, nearly periodic signals.
  • For contact electrograms, recompose from slope-weighted sine wavelets before taking the phase.

Phase singularities

Walk around a small closed loop on a phase map and add up the changes in phase, each wrapped into \((-\pi, \pi]\). The total is a whole multiple of \(2\pi\):

\[ n = \frac{1}{2\pi} \oint \nabla\varphi \cdot d\boldsymbol{\ell} = \frac{1}{2\pi} \sum_{k} \operatorname{wrap}\!\left(\varphi_{k+1} - \varphi_k\right) \tag{6}\]

\(n\) is the topological charge. It is zero for any loop that does not enclose a phase singularity (PS) and \(\pm 1\) around one, the sign giving the direction of rotation (Bray et al. 2001; Bray and Wikswo 2002). A spiral wave has one PS at its tip; a plane wave has none.

On a real catheter the loop is a ring of electrodes, and noise or interpolation can make the sum come out at \(\pm 2\pi\) by chance. In high-density epicardial AF recordings, at least eight electrodes in a ring were needed to avoid frequent false-positive PS due to chance (Kuklik et al. 2017).

Key takeaways

  • Topological charge counts net phase winding around a loop: \(\pm1\) at a PS, 0 elsewhere.
  • Few electrodes per ring means false singularities; test any detector on data without rotors.

How fine must the electrodes be?

A train of wavefronts moving at speed \(v\) and repeating every \(\mathrm{CL}\) has its fronts a distance

\[ \lambda_s = v \cdot \mathrm{CL} \tag{7}\]

apart: the spatial wavelength. The tests measure it in a paced cable model. Simulations show that the electrode spacing needed to identify rotors and focal sources correctly is a linear function of this wavelength, and that rotor localization errors are larger with electrograms than with action potentials (Roney et al. 2017). A catheter that is coarse compared with \(\lambda_s\) cannot tell a rotor from other patterns.

Key takeaways

  • Spatial wavelength = conduction velocity × cycle length.
  • Required electrode spacing scales with that wavelength; coarse sampling creates and hides drivers.

Clinical correlate

NotePending clinical review

Planned in the outline. Requires sign-off by an EP physician.

Visual summary

Planned in the outline.

How experts think

Planned in the outline.

Cheat sheet

Planned in the outline.

Interview questions

Planned in the outline.

Code lab

Planned in the outline.

Self-check

Planned in the outline.

Literature figures

Planned in the outline.

Further reading

Planned in the outline.

References

Bray, M A, S F Lin, R R Aliev, et al. 2001. “Experimental and Theoretical Analysis of Phase Singularity Dynamics in Cardiac Tissue.” Journal of Cardiovascular Electrophysiology 12 (6): 716–22. https://doi.org/10.1046/j.1540-8167.2001.00716.x.
Bray, Mark-Anthony, and John P Wikswo. 2002. “Use of Topological Charge to Determine Filament Location and Dynamics in a Numerical Model of Scroll Wave Activity.” IEEE Transactions on Biomedical Engineering 49 (10): 1086–93. https://doi.org/10.1109/TBME.2002.803516.
Everett, T H, L C Kok, R H Vaughn, et al. 2001. “Frequency Domain Algorithm for Quantifying Atrial Fibrillation Organization to Increase Defibrillation Efficacy.” IEEE Transactions on Biomedical Engineering 48 (9): 969–78. https://doi.org/10.1109/10.942586.
Gray, R A, A M Pertsov, and J Jalife. 1998. “Spatial and Temporal Organization During Cardiac Fibrillation.” Nature 392 (6671): 75–78. https://doi.org/10.1038/32164.
Kuklik, Pawel, Stef Zeemering, Arne van Hunnik, et al. 2017. “Identification of Rotors During Human Atrial Fibrillation Using Contact Mapping and Phase Singularity Detection: Technical Considerations.” IEEE Transactions on Biomedical Engineering 64 (2): 310–18. https://doi.org/10.1109/TBME.2016.2554660.
Kuklik, Pawel, Stef Zeemering, Bart Maesen, et al. 2015. “Reconstruction of Instantaneous Phase of Unipolar Atrial Contact Electrogram Using a Concept of Sinusoidal Recomposition and Hilbert Transform.” IEEE Transactions on Biomedical Engineering 62 (1): 296–302. https://doi.org/10.1109/TBME.2014.2350029.
Nademanee, Koonlawee, John McKenzie, Erol Kosar, et al. 2004. “A New Approach for Catheter Ablation of Atrial Fibrillation: Mapping of the Electrophysiologic Substrate.” Journal of the American College of Cardiology 43 (11): 2044–53. https://doi.org/10.1016/j.jacc.2003.12.054.
Ng, Jason, and Jeffrey J Goldberger. 2007. “Understanding and Interpreting Dominant Frequency Analysis of AF Electrograms.” Journal of Cardiovascular Electrophysiology 18 (6): 680–85. https://doi.org/10.1111/j.1540-8167.2007.00832.x.
Ng, Jason, Alan H Kadish, and Jeffrey J Goldberger. 2007. “Technical Considerations for Dominant Frequency Analysis.” Journal of Cardiovascular Electrophysiology 18 (7): 757–64. https://doi.org/10.1111/j.1540-8167.2007.00810.x.
Roney, Caroline H, Chris D Cantwell, Jason D Bayer, et al. 2017. “Spatial Resolution Requirements for Accurate Identification of Drivers of Atrial Fibrillation.” Circulation: Arrhythmia and Electrophysiology 10 (5): e004899. https://doi.org/10.1161/CIRCEP.116.004899.
Sanders, Prashanthan, Omer Berenfeld, Mélèze Hocini, et al. 2005. “Spectral Analysis Identifies Sites of High-Frequency Activity Maintaining Atrial Fibrillation in Humans.” Circulation 112 (6): 789–97. https://doi.org/10.1161/CIRCULATIONAHA.104.517011.