m28 · Pulsed field ablation: electroporation and what it changes for mapping

Warning

Education only. Not for diagnosis, treatment or any clinical decision.

Note

Draft, stage 2 of 6. The derivations are written, and each equation is checked by tests in tests/test_m28_pfa.py. Demos, code lab, questions and the remaining sections follow in later stages (plan: docs/outlines/m28-outline.md). Nothing here has had expert review yet. This module explains physics; it does not recommend waveforms, catheters or workflows for patients.

In 60 seconds

Pulsed field ablation (PFA) delivers very short, high-voltage pulses. The electric field they set up in tissue charges every cell membrane it crosses. Where the induced membrane voltage is large enough, the membrane becomes permeable: briefly if the field is moderate, for good if it is strong. The lesion is therefore shaped by where the field exceeds a threshold, not by where tissue gets hot. Two consequences matter for mapping. The field map, not a heat map, predicts the lesion. And a shell of tissue around the lesion can stop conducting for a while and then recover, so a map made right after an application can differ from one made later.

From the field

Planned in the outline (docs/outlines/m28-outline.md); written in a later stage.

Derivations

Symbols used throughout. \(E\) magnitude of the electric field (V/m). \(R\) cell radius (m). \(\theta\) angle between the field and a point on the membrane. \(\Delta\Psi_m\) induced transmembrane voltage (V). \(C_m\) membrane capacitance per area (F/m²). \(\rho_i\), \(\rho_e\) intracellular and extracellular resistivity (Ω·m), with \(\sigma = 1/\rho\). \(V\) electrode voltage (V), \(r_0\) electrode radius (m), \(r\) distance from the electrode centre (m). \(E_{\text{th}}\) a threshold field (V/m).

How a field charges a membrane

Short, high-voltage pulses make cell membranes more permeable for a time. Pores forming in the lipid bilayer are the widely recognized mechanism; changes to individual membrane lipids and proteins also contribute, which is why some authors separate the terms electroporation and electropermeabilization (Kotnik et al. 2019). Permeability follows the voltage the field induces across the membrane, so that is what we compute.

Model a cell as a sphere of radius \(R\) whose membrane does not conduct, in a uniform field \(E\). Outside the cell the potential must satisfy Laplace’s equation, carry no current into the membrane, and look like the applied field far away. The solution is \(\varphi_{\text{out}} = -E\left(r + \tfrac{R^3}{2r^2}\right)\cos\theta\). Inside, no current flows, so the potential is uniform; call it zero. The membrane voltage is the difference at \(r = R\):

\[ \Delta\Psi_m(\theta) = \tfrac{3}{2}\, E\, R \cos\theta \tag{1}\]

This is Schwan’s equation (Kotnik and Miklavčič 2000). The voltage is largest at the two poles that face the electrodes and zero at the equator, so pores form first at the poles. Larger cells feel a larger voltage in the same field. Kotnik and Miklavčič generalized the result to spheroidal cells, for which the induced voltage also depends on how the cell is oriented in the field (Kotnik and Miklavčič 2000); heart cells are elongated, so their orientation to the field matters.

How fast the membrane charges. A real interior and exterior conduct, and the membrane is a capacitor. Keep the same \(\cos\theta\) shape inside and outside, require the radial current to be continuous across the membrane and equal to \(C_m\, d\Delta\Psi_m/dt\). The membrane voltage then obeys a first-order equation, and after a step in field at \(t = 0\):

\[ \Delta\Psi_m(\theta, t) = \tfrac{3}{2}\, E\, R \cos\theta \left(1 - e^{-t/\tau_m}\right), \qquad \tau_m = R\, C_m \left(\rho_i + \tfrac{\rho_e}{2}\right) \tag{2}\]

A pulse much shorter than \(\tau_m\) charges the membrane only partly; a pulse much longer than \(\tau_m\) reaches the steady value of Equation 1. Values of \(C_m\), \(\rho_i\) and \(\rho_e\) for heart cells: TODO(source).

Key takeaways

  • A field \(E\) induces \(\tfrac32 E R\cos\theta\) across a spherical membrane: largest at the poles, larger for larger cells.
  • The membrane charges with time constant \(R C_m(\rho_i + \rho_e/2)\); pulse width matters.

The field around an electrode and the lesion edge

Model the electrode as a sphere of radius \(r_0\) held at voltage \(V\), with the return far away, in uniform tissue. The potential solves Laplace’s equation with spherical symmetry:

\[ \varphi(r) = V\,\frac{r_0}{r}, \qquad E(r) = V\,\frac{r_0}{r^2} \tag{3}\]

If cells die wherever the field exceeds a threshold \(E_{\text{th}}\), the lesion edge is the iso-field surface \(E(r_L) = E_{\text{th}}\):

\[ r_L = \sqrt{\frac{V\, r_0}{E_{\text{th}}}} \tag{4}\]

Quadrupling the voltage doubles the lesion radius. Compare RF (m27): there the edge is an isotherm set by heat that spreads over time; here it is set by the instantaneous field, so it does not grow with time in the same way.

The threshold is not one number. Electroporation is governed mainly by the local field and by lethal thresholds that depend on the waveform: pulse shape, width, number and duty cycle (Anić et al. 2026). Bench studies show how much the values move. In a cell-suspension model with monophasic pulses, cardiomyocyte and neuronal lines had less than 20 % viability at 1000 V/cm with at least 30 pulses, and cardiac fibroblasts needed 1250 V/cm with at least 50 pulses (Avazzadeh et al. 2021). In a culture model, neurons were more susceptible than cardiomyocytes at 1250 V/cm, and a significant share of cardiomyocytes recovered by 24 hours (Avazzadeh et al. 2022). These are bench values; tissue thresholds in the beating heart: TODO(source).

Reversible and irreversible zones. The increase in permeability can be transient (Kotnik et al. 2019), and in suspension some cell membranes resealed within hours of the pulses (Avazzadeh et al. 2021). If membranes open reversibly above \(E_{\text{rev}}\) and irreversibly above \(E_{\text{irr}} > E_{\text{rev}}\), applying Equation 4 twice gives two shells:

\[ r_{\text{rev}} = \sqrt{\frac{V\, r_0}{E_{\text{rev}}}}, \qquad r_{\text{irr}} = \sqrt{\frac{V\, r_0}{E_{\text{irr}}}}, \qquad \frac{r_{\text{rev}}}{r_{\text{irr}}} = \sqrt{\frac{E_{\text{irr}}}{E_{\text{rev}}}} \tag{5}\]

Tissue between the two radii may stop conducting for a while and then recover. For mapping this predicts a “core” that stays silent and a “halo” that comes back. One clinical study with a multi-spline catheter reported exactly that pattern on repeated unipolar and bipolar voltage maps: a stable core of low unipolar voltage and a surrounding halo that recovered (Na et al. 2026).

Key takeaways

  • Around a spherical electrode \(E = V r_0/r^2\); an iso-field lesion has radius \(\sqrt{V r_0/E_{\text{th}}}\).
  • Thresholds depend on cell type and waveform; bench values are not tissue values.
  • A reversible shell lies outside the lesion, \(\sqrt{E_{\text{irr}}/E_{\text{rev}}}\) times wider: the acute map can overstate the lasting lesion.

Two electrodes

A PFA catheter can pass current between its own electrodes (bipolar) or from the catheter to a distant patch (monopolar); the choice changes the field distribution (Anić et al. 2026). For two small spheres of radius \(r_0\) at positions \(\mathbf{x}_1\) and \(\mathbf{x}_2\), held at \(V_1\) and \(V_2\) and far apart compared with their size, the potentials add:

\[ \varphi(\mathbf{x}) \approx V_1\,\frac{r_0}{\lvert \mathbf{x} - \mathbf{x}_1\rvert} + V_2\,\frac{r_0}{\lvert \mathbf{x} - \mathbf{x}_2\rvert} \tag{6}\]

The sum is not exact: on the surface of the first sphere it adds about \(V_2\, r_0 / d\) to the voltage that sphere should have, where \(d\) is the distance between centres. The error shrinks as the electrodes move apart. With \(V_2 = -V_1\) the field is strongest between the electrodes, so the lesion is shaped by the electrode layout as well as by voltage.

Key takeaways

  • Superposition of single-electrode fields is a good model for small, well-separated electrodes.
  • Electrode layout shapes the iso-field contour, and therefore the lesion.

Heating is small, but not zero

Current through tissue still heats it. During a pulse the power density is \(\sigma E^2\). Ignoring conduction gives an upper bound on the rise in temperature over a pulse train with total on-time \(t_{\text{on}}\), for tissue with volumetric heat capacity \(\rho c\) (J/(m³·K)):

\[ \Delta T \le \frac{\sigma\, E^2\, t_{\text{on}}}{\rho\, c} \tag{7}\]

Conduction and blood flow can only lower the actual rise. Short pulses keep \(t_{\text{on}}\) small, which is why PFA is described as largely non-thermal; high-voltage pulses can still cause Joule heating, electrochemical reactions and bubbles, which waveform and electrode design must control (Anić et al. 2026).

Key takeaways

  • The adiabatic bound \(\sigma E^2 t_{\text{on}}/(\rho c)\) grows with the square of the field and with total on-time.
  • “Largely non-thermal” is a statement about \(t_{\text{on}}\), not about the absence of heating.

Clinical correlate

NotePending clinical review

Planned in the outline. Requires sign-off by an EP physician.

Visual summary

Planned in the outline.

How experts think

Planned in the outline.

Cheat sheet

Planned in the outline.

Interview questions

Planned in the outline.

Code lab

Planned in the outline.

Self-check

Planned in the outline.

Literature figures

Planned in the outline.

Further reading

Planned in the outline.

References

Anić, Ante, Antoni Ivorra, Brendan Koop, et al. 2026. “Pulsed Field Ablation: From Concepts to Technology.” Europace 28 (Supplement_2): ii18–31. https://doi.org/10.1093/europace/euag205.
Avazzadeh, Sahar, Mahshid H Dehkordi, Peter Owens, et al. 2022. “Establishing Electroporation Thresholds for Targeted Cell Specific Cardiac Ablation in a 2D Culture Model.” Journal of Cardiovascular Electrophysiology 33 (9): 2050–61. https://doi.org/10.1111/jce.15641.
Avazzadeh, Sahar, Barry O’Brien, Ken Coffey, et al. 2021. “Establishing Irreversible Electroporation Electric Field Potential Threshold in a Suspension in Vitro Model for Cardiac and Neuronal Cells.” Journal of Clinical Medicine 10 (22): 5443. https://doi.org/10.3390/jcm10225443.
Kotnik, Tadej, and Damijan Miklavčič. 2000. “Analytical Description of Transmembrane Voltage Induced by Electric Fields on Spheroidal Cells.” Biophysical Journal 79 (2): 670–79. https://doi.org/10.1016/S0006-3495(00)76325-9.
Kotnik, Tadej, Lea Rems, Mounir Tarek, et al. 2019. “Membrane Electroporation and Electropermeabilization: Mechanisms and Models.” Annual Review of Biophysics 48: 63–91. https://doi.org/10.1146/annurev-biophys-052118-115451.
Na, Jonathan, Monisha Krishna Murthy, Helena Lopez-Martinez, et al. 2026. “Unipolar Voltage Mapping to Predict Lesion Durability During Pulsed Field Ablation.” Heart Rhythm 23 (8): e1397–407. https://doi.org/10.1016/j.hrthm.2026.01.025.